Biotage - TurboVap 96
Microprocessor-controlled evaporation system for simultaneous, automated concentration of multiple samples with unattended operation.
Applications
TurboVap® 96 has two independent compartments that accommodate 1 mL or 2 mL standard or deep well plates. Each compartment has independent control of gas flow and temperature. The flow rate of the gas is determined by the volume of solvent in the plate and the temperature values can be set with the help of an evaporation guide supplied. Nozzles are easy to clean and require minimal maintenance. In genomic and proteomic applications, the evaporation of DMSO from plates is a common application. As is drying HPLC aqueous solvent mixtures of methanol or acetonitrile. TurboVap® 96 complements automated 96-well SPE workstations that are widely used for drugs of abuse determination and other pharmaceutical SPE methods.
Active Questions & AnswersAsk a Question
Front Panel
Vortex Service ProvidersView All (9)
Documents & Manuals
There are no Documents or Manuals available.
Features of TurboVap 96
Features
- Holds two standard microplates or deep-well plates
- Individual plate control for gas flow, temperature and timed endpoint
- Touch Pad Control panel retains the previous settings for faster start-ups
- TFA, DMSO and DMF compatible
- Patented gas vortex technology keeps the needles away from the solvent eliminating carry-over concerns
- Improved safety with automatic gas, time and pressure shutoffs
- No hood space required
General Specifications
Depth | 12 in |
Height | 12 in |
Width | 24 in |
Power Requirements | 100/120VAC |
Weight | 40.5 lbs |
Evaporator Type | Vortex |
Additional Specifications
Power requirements 100/120V, 50/60 Hz, 10Amps, 220/240V, 50/60 Hz, 5Amps
Part numbers P/N 103263 - 100/120V, 50/60 Hz, P/N 103264 - 220/240V 60/60 Hz
Dimensions 24\" X 12\" x 12”
Capacity 2 standard 96 well microplates or deep-well plates
Format 8 X 12 well microplates and deep-well plates
Gas flow 0-90SCFH
Exhaust 25 CFM blower
Temperature range Ambient to 80°
Technology Gas vortex shearing technology
Number of samples 2 x 96 well plates, standard or deep with adapters
Timer range 1 to 99 minutes
Max. sample volume <2 mL
Final endpoint volumes -
Heaters Two internal individually controlled heaters. Each adjustable fr
Gas supply requirements Minimum inlet pressure 30 psi (2.1 bar) Maximum inlet pressure 80 psi (5.5 bar)
Range 5–99 cfm
Fuses 100/120VAC model: 6A , T250V 220/240 VAC model: 3A, T250V
Dimensions (WxDxH) 30.5 x 30.5 x 61 cm 12\" x 12\" x 24\"
Weight 18.4 kg/40.5 lbs
Certifications 2006/95/EC Low Voltage Directive 2004/108/EC EMC Directive 93/68