SP Scientific - ThermoJet ES Precision Temperature Cycling System
ThermoJet Environment Simulator is a precision-controlled temperature forcing system that delivers an airstream of up...
ThermoJet provides temperature transition speeds unmatched in the industry, and can sustain air cooling temperatures of –80°C at 20 SCFM. It is made with the proven and reliable Dual Cool cascade refrigeration system and does not require complex and difficult blended gases.
The ThermoJet provides the smallest footprint for a floor model air stream temperature cycling system
Active Questions & AnswersAsk a Question
There are no current Discussions
Temperature Controller Service ProvidersView All (8)
Documents & ManualsView All Documents
Features of ThermoJet ES Precision Temperature Cycling System
- Temperature range of +225°C to -80°C * High Temperature option of +350°C * Delivers 5 to 20 standard cubic feet of air/minute * Temperature transition from +200°C to -65°C in less than 15 seconds -80°C to +200°C in less than 15 seconds * Reliable and robust Dual Cool cascade refrigeration system does not require difficult blended gases * Small footprint allows maximization of space usage * External Device Under Test thermocouple for controlling temperature accuracy at the device * Configured with no holding arm or mechanical arm for easy air stream placement * Dry air purge to help protect test site * Selection of shrouds and boxes to meet your application * Built in air dryer with oil, water and particulate filters * Enhanced BTU/watt removal capabilities provided by higher air flow rate * Effortless change temperature and air flow set points
General Specifications
Temperature Range | -80 to 350 °C |
Temperature Accuracy | 1 °C - Factory calibrated with NIST traceable device |
Additional Specifications
Setability:
0.1°C
Indication/Resolution:
0.1°C
Stability:
±0.5°C at discharge nozzle
Sensors:
Special limits of error type-T thermocouples
Calibration:
Software controlled
Max. Transition Rates:
+125°C to -55°C in <10s / -55°C to +125°C in <10s
+200°C to -65°C in <15s / -80°C to +200°C in <15s
Stabilization Rate3:
+125°C to -55°C in <12s
-55°C to +125°C in <4s
Air Filtering/Drying
Coalescent:
Water, Oil, and Particle Filtering
Heatless Regenerative:
-85°C dew point
Air Flow Specifications
Input:
25 SCFM @ 80-125 psig, dew point 10°C
(11.8 l/s @ 5.5 - 8.6 bar)
Output:
5-20 SCFM (2.3 - 9.4 l/s)
Factory calibrated with NIST traceable device
Setability:
1 SCFM (0.4 l/s)
Physical Characteristics
Footprint:
21" wide x 27" deep (54 cm x 69 cm)
Arm Reach:
56" radius (140 cm) max.
Rotation:
350°
Work Surface Height:
38" (100 cm)
Weight:
425 lbs (193 kg)
Mobility:
4 full swivel, lockable casters
DB Rating4:
@ 65 dB(A)
System Controller Features
Type:
Microprocessor PLC-based PID
Program:
Save and load 48 programs of 8 segments each
Safeties:
Low flow, over temp, EMo interlock
Status Indication:
Air input, air temp, air/DUT control,
OK, run/idle, purge, defrost
Available Limits:
Max. ramp rate, min./max. air temperatures
Ramp Rates:
1°C/min. incremental up to 999°C
Remote Information:
IEEE, RS232 Communication
System Requirements
Temperature:
+12°C to +25°C
Humidity:
0% to 65% RH