Waters - 717
The Waters 717 Autosampler injects samples into high performance liquid chromatography system. The Waters 717...
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Autosamplers Service ProvidersView All (1)
Documents & ManualsView All Documents
Features
There are no Features available.
General Specifications
| Cross Contamination | 0.1 % |
| Maximum Pressure | 4500 psi maximum continuous operating pressure. 5000 psig maximum surge for 30 seconds. |
| Standard Injection Range | 0.1 to 2000 μL |
Additional Specifications
Linearity (1 to 100 uL): 0.99 Coefficient of deviation
Bandspreading: 10% Contribution for all solvent systems
Precision (typical, 5 to 50 uL): 0.5 % RSD
Wash Flow, Solvent Rate: 0.4 mL per injection
Minimum Sample Required: 7 uL using low volume inserts
[u]Mechanical Specifications[/u]
Wetted Surface Materials: 316 Stainless Steel, Teflon, Glass, Fluoraloy-08 R, UHMWPE
Number of Sample Vials: 1 to 96
Number of Injections: Programmable 1 to 99 injections per vial
Allowable Solvents: All standard LC solvents
Inlet Tubing to Injector: 0.040 I.D. stainless steel
Outlet Tubing from Injector: 0.009 ID stainless steel
Total Solvent Volume: 836 uL
[u]Environmental Specifications[/u]
Operating Temperature Range: 4-40 °C
Relative Humidity: 80%, non-condensing
Storage Temperature Range: - 40 to 70 °C