Labconco - Protector Premier
Protector Premier Laboratory Hoods
Protector Premier Laboratory Hoods are suitable for general chemistry applications. Explosion-proof models are available for applications involving high concentrations of flammable solvents.
Protector Premier Laboratory Hoods are benchtop design and are offered in 4', 5', 6' and 8' widths so their dimensions exactly match standard laboratory casework or accessory Protector Base Cabinets and Stands. Choose from hoods designed to be used with a remotely-located blower or hoods that include a built-in blower.
Active Questions & AnswersAsk a Question
There are no current Discussions
Fume Hoods Service ProvidersView All (9)
Documents & ManualsView All Documents
Features of Protector Premier
- By-pass airflow design
- Aerodynamic air foil allows air to sweep the work surface while Clean-Sweep™ openings pull inflow air from under the air foil.
- Glacier white, dry powder epoxy-coated steel exterior
- One-piece molded fiberglass liner and pre-set baffle(s) with flame spread less than 25 per ASTM E-84
- 3/16" thick tempered safety glass vertical-rising sash(es) with epoxy-coated aluminum sash handle
- Removable front and side panels and front access panels for access to plumbing and electrical wiring
- Pre-wired instant-start T8 fluorescent lighting, light and blower switches (non-explosion-proof models only)
- Factory-prepared for up to eight service fixtures including two gooseneck faucets
- Factory-prepared for four electrical duplex receptacles
- ETL listed (115 volt models)
- CE conformity (230 volt models)
- 4', 5', 6', 8' widths
- International electrical configurations
- Built-in blowers
- Explosion-proof lighting and blower
- Remote blowers
- Ductwork
- SpillStopper™ Work Surfaces
- Base Cabinets
- Adjustable Height Base Stands
- Service Fixture Kits
- Electrical Duplex Kits
- Airflow monitors
- Distillation grids
- Sash stops
- Snuffer Fire Extinguisher
- Ceiling enclosures
- Rear finish panels
- Auxiliary-air plenums
General Specifications
There are no General Specifications available.
Additional Specifications
Product Subcategory: General Chemistry
Nominal Width: 4'
Electrical Requirements: International
Blower Requirements: Remote blower required
Air Usage/Static Pressure with Sash Full Open: At 100 fpm, exhausts 730 CFM at 0.17" H2O static pressure
Air Usage/Static Pressure with Sash Open 18": At 100 fpm, exhausts 470 CFM at 0.07" H2O static pressure
Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805, CE Conformity Marking (230 volt models)
Style: Benchtop
Other: 4' Protector Premier Laboratory Hood, by-pass airflow design, chemical-resistant molded one-piece fiberglass liner, preset baffle, 3/16" thick tempered safety glass vertical-rising counterbalanced sash, ergonomic air foil with Clean-Sweep openings, 12.81" ID exhaust connection, front access panels, ADA-compliant light and blower switches, glacier white dry powder epoxy-coated steel exterior. Contact Labconco for explosion-proof models.
Series Modals
Labconco - 4ft Protector Premier
[+] View Model Specifications| Depth | 33.2 in |
| Height | 59 in |
| Width | 48 in |
| Power Requirements | Available in Domestic and International |
| Lighting | T8 fluorescent |
| Face Velocity (FPM) | 100 |
| Fume Hood Type | Conventional Ducted |
| Additional Model Specifications | Product Subcategory: General Chemistry
Style: Benchtop Nominal Width: 4' Blower Requirements: Remote blower required Air Usage/Static Pressure with Sash Full Open: At 100 fpm, exhausts 730 CFM at 0.17" H2O static pressure Air Usage/Static Pressure with Sash Open 18": At 100 fpm, exhausts 470 CFM at 0.07" H2O static pressure Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805, CE Conformity Marking (230 volt models) Other: By-pass airflow design, chemical-resistant molded one-piece fiberglass liner, preset baffle, 3/16" thick tempered safety glass vertical-rising counterbalanced sash, ergonomic air foil with Clean-Sweep openings, 12.81" ID exhaust connection, front access panels, ADA-compliant light and blower switches, glacier white dry powder epoxy-coated steel exterior. Contact Labconco for explosion-proof models. Power Supply: Models 4871500 & 4871600: 230 V, 50 Hz, International Models 4880100 & 4880400: 115 V, 60 Hz, Domestic |
Labconco - 4ft Protector Premier with built-in exhaust blower
[+] View Model Specifications| Airflow (CFM) | 730 CFM |
| Depth | 33.2 in |
| Height | 65.63 in |
| Width | 48 in |
| Power Requirements | Available in Domestic and International |
| Lighting | T8 fluorescent |
| Face Velocity (FPM) | 100 |
| Fume Hood Type | Conventional Ducted |
| Additional Model Specifications | Product Subcategory: General Chemistry
Style: Benchtop Nominal Width: 4' Blower Requirements: Integral blower included Air Usage/Static Pressure with Sash Full Open: At 100 fpm, exhausts 730 CFM at 0.17" H2O static pressure Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805, CE Conformity Marking (230 volt models) Other: By-pass airflow design, chemical-resistant molded one-piece fiberglass liner, preset baffle, 3/16" thick tempered safety glass vertical-rising counterbalanced sash, ergonomic air foil with Clean-Sweep openings, 12.81" ID exhaust connection, front access panels, ADA-compliant light and blower switches, glacier white dry powder epoxy-coated steel exterior. Contact Labconco for explosion-proof models. Power Supply: Models 4871300 & 4871400: 230 V, 50 Hz, International Models 4880000, 4880200 & 4880300: 115 V, 60 Hz, Domestic Model 4880200: Explosion-proof |
Labconco - 5ft Protector Premier
[+] View Model Specifications| Depth | 33.2 in |
| Height | 59 in |
| Width | 60 in |
| Power Requirements | Available in Domestic and International |
| Lighting | T8 fluorescent |
| Face Velocity (FPM) | 100 |
| Fume Hood Type | Conventional Ducted |
| Additional Model Specifications | Product Subcategory: General Chemistry
Style: Benchtop Nominal Width: 5' Blower Requirements: Remote blower required Air Usage/Static Pressure with Sash Full Open: At 100 fpm, exhausts 960 CFM at 0.24" H2O static pressure Air Usage/Static Pressure with Sash Open 18": At 100 fpm, exhausts 610 CFM at 0.11" H2O static pressure Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805, CE Conformity Marking (230 volt models) Other: By-pass airflow design, chemical-resistant molded one-piece fiberglass liner, preset baffle, 3/16" thick tempered safety glass vertical-rising counterbalanced sash, ergonomic air foil with Clean-Sweep openings, 12.81" ID exhaust connection, front access panels, ADA-compliant light and blower switches, glacier white dry powder epoxy-coated steel exterior. Contact Labconco for explosion-proof models. Power Supply: Models 6071500 & 6071600: 230 V, 50 Hz, International Models 6080100 & 6080400: 115 V, 60 Hz, Domestic Model 4880200: Explosion-proof |
Labconco - 5ft Protector Premier with built-in exhaust blower
[+] View Model Specifications| Airflow (CFM) | 960 CFM |
| Depth | 33.2 in |
| Height | 65.63 in |
| Width | 60 in |
| Power Requirements | Available in Domestic and International |
| Lighting | T8 fluorescent |
| Face Velocity (FPM) | 100 FPM |
| Fume Hood Type | Conventional Ducted |
| Additional Model Specifications | Product Subcategory: General Chemistry
Style: Benchtop Nominal Width: 5' Blower Requirements: Integral blower included Air Usage/Static Pressure with Sash Full Open: At 100 fpm, exhausts 960 CFM at 0.12" H2O maximum external static pressure Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805, CE Conformity Marking (230 volt models) Other: By-pass airflow design, chemical-resistant molded one-piece fiberglass liner, preset baffle, 3/16" thick tempered safety glass vertical-rising counterbalanced sash, ergonomic air foil with Clean-Sweep openings, 12.81" ID exhaust connection, front access panels, ADA-compliant light and blower switches, glacier white dry powder epoxy-coated steel exterior. Contact Labconco for explosion-proof models. Power Supply: Models 6071300 & 6071400: 230 V, 50 Hz, International Models 6080000, 6080200 & 6080300: 115 V, 60 Hz, Domestic Model 6080200: Explosion-proof |
Labconco - 6ft Protector Premier
[+] View Model Specifications| Depth | 33.2 in |
| Height | 59 in |
| Width | 72 in |
| Power Requirements | Available in Domestic and International |
| Lighting | T8 fluorescent |
| Face Velocity (FPM) | 100 FPM |
| Fume Hood Type | Conventional Ducted |
| Additional Model Specifications | Product Subcategory: General Chemistry
Style: Benchtop Nominal Width: 6' Blower Requirements: Remote blower required Air Usage/Static Pressure with Sash Full Open: At 100 fpm, exhausts 1180 CFM at 0.31" H2O static pressure Air Usage/Static Pressure with Sash Open 18": At 100 fpm, exhausts 750 CFM at 0.14" H2O static pressure Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805, CE Conformity Marking (230 volt models) Other: By-pass airflow design, chemical-resistant molded one-piece fiberglass liner, preset baffle, 3/16" thick tempered safety glass vertical-rising counterbalanced sash, ergonomic air foil with Clean-Sweep openings, 12.81" ID exhaust connection, front access panels, ADA-compliant light and blower switches, glacier white dry powder epoxy-coated steel exterior. Contact Labconco for explosion-proof models. Power Supply: Models 7271500 & 7271600: 230 V, 50 Hz, International Models 7280100 & 7280400: 115 V, 60 Hz, Domestic |
Labconco - 6ft Protector Premier with built-in exhaust blower
[+] View Model Specifications| Airflow (CFM) | 1180 CFM |
| Depth | 33.2 in |
| Height | 65.63 in |
| Width | 72 in |
| Power Requirements | Available in Domestic and International |
| Lighting | T8 fluorescent |
| Face Velocity (FPM) | 100 FPM |
| Fume Hood Type | Conventional Ducted |
| Additional Model Specifications | Product Subcategory: General Chemistry
Style: Benchtop Nominal Width: 6' Blower Requirements: Integral blower included Air Usage/Static Pressure with Sash Full Open: At 100 fpm, exhausts 1180 CFM at 0.17" H2O maximum external static pressure Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805, CE Conformity Marking (230 volt models) Other: By-pass airflow design, chemical-resistant molded one-piece fiberglass liner, preset baffle, 3/16" thick tempered safety glass vertical-rising counterbalanced sash, ergonomic air foil with Clean-Sweep openings, 12.81" ID exhaust connection, front access panels, ADA-compliant light and blower switches, glacier white dry powder epoxy-coated steel exterior. Contact Labconco for explosion-proof models. Power Supply: Models 7271300 & 7271400: 230 V, 50 Hz, International Models 7280000, 7280200 & 7280300: 115 V, 60 Hz, Domestic Model 7280200: Explosion-proof |
Labconco - 8ft Protector Premier
[+] View Model Specifications| Depth | 33.2 in |
| Height | 59 in |
| Width | 96 in |
| Power Requirements | Available in Domestic and International |
| Lighting | T8 fluorescent |
| Face Velocity (FPM) | 100 FPM |
| Fume Hood Type | Conventional Ducted |
| Additional Model Specifications | Product Subcategory: General Chemistry
Style: Benchtop Nominal Width: 8' Blower Requirements: Remote blower required Air Usage/Static Pressure with Sash Full Open: At 100 fpm, exhausts 1660 CFM at 0.22" H2O static pressure Air Usage/Static Pressure with Sash Open 18": At 100 fpm, exhausts 1060 CFM at 0.08" H2O static pressure Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805, CE Conformity Marking (230 volt models) Other: By-pass airflow design, chemical-resistant molded one-piece fiberglass liner, preset baffle, 3/16" thick tempered safety glass vertical-rising counterbalanced sash, ergonomic air foil with Clean-Sweep openings, 12.81" ID exhaust connection, front access panels, ADA-compliant light and blower switches, glacier white dry powder epoxy-coated steel exterior. Contact Labconco for explosion-proof models. Power Supply: Model 9674500: 230 V, 50 Hz, International Model 9683000: 115 V, 60 Hz, Domestic |