Features of MCO-230AICUVH
  • Uniform Consistent Environment
  • Rapid CO2 Recovery with IR2 Sensors
  • New LCD Touch Screen for Improved Usability
  • Easier to Clean and Maintain
  • Condensation Management System for improved Humidity
  • 3 Hour H2O2 Decontamination
  • Active Background Decontamination Control with SafeCell UV and InCu-saFe®
  • USB Log Function for Data Transfer and Storage
General Specifications
Depth 28.7 in
Height35.6 in
Internal Depth 20.6 in
Internal Height 27.6 in
Width 30.3 in
Internal Width 25.3 in
Chamber Volume230 L
CO2 Range0 to 20 %
Weight90 kg
Additional Specifications

DATA ACQUISITION   
Data acquisition system (MTR-5000)-Optional   
OUTER DOOR   
Painted Steel   
SHELVES   
4 trays made of InCuSaFe stainless steel containing copper W24.4 inch x D17.7 inch (W620 mm x D450 mm)  
Maximum load: 7 kg/tray   
ACCESS PORT   
Inner Diameter: 30 mm (1.18 inch), On the back side   
INSULATION   
Extruded polystyrene foam   
AIR FILTER   
0.3 um, Efficiency: 99.97 % or higher   
UV LAMP   
4W x 1 (ozone-free emission)    
TEMPERATURE CONTROL RANGE   
Ambient temperature +º5C to 50ºC (ambient temperature: 5ºC to 35ºC)   
TEMPERATURE UNIFORMITY   
±0.25ºC (ambient temperature: 23ºC, setting: 37ºC, CO2: 5 %, no load)   
MAXIMUM POWER CONSUMPTION   
Max. 430 W    
CO2 CONTROLLER   
PID Control System   
CO2 GAS CONNECTION   
4 mm to 6 mm (0.157 inch to 0.36 inch) diameter tube can be connected   
CO2 GAS PRESSURE   
0.03 MPa(G) (0.3kgf/cm2(G),4.3 psi(G))   
CO2 VARIATION   
±0.15% (ambient temperature: 23ºC, setting: 37ºC, CO2: 5%, no load)   
CO2 SENSOR TYPE   
Dual Infrared Sensor   
DATA OUTPUT   
USB   
DECONTAMINATION METHODS   
H2O2 Decontamination - Standard 24 hr full chamber UV decontamination SafeCell UV inCu-saFe Active Background Decontamination control   
CONTAMINATION CONTROLS   
SafeCell UV inCu-saFe Active Background Decontamination control   
DECONTAMINATION CYCLE TIME   
< 3.0 Hours   
DISPLAY   
Color LCD touch panel   
HEATING SYSTEM   
DHA System (Heater Jacket + Air Jacket System)   
HUMIDIFYING SYSTEM   
Natural evaporation with humidifying pan,  
95 %R.H.±5 %R.H.   
INTERIOR SURFACES   
InCu-saFe® Stainless steel containing copper