PHCBi - MCO-5M/MCO-5MUV
CytoGROW Compact Multigas Incubators
Active Questions & AnswersAsk a Question
There are no current Discussions
CO2 Incubator Service ProvidersView All (28)
Documents & ManualsView All Documents
Features of MCO-5M/MCO-5MUV
MCO-5M
- Compact Design and Stackable
- InCu-saFe® Copper Stainless Steel Alloy
- SafeCell UV Protection (Optional)
- Direct Heat and Air Jacket System
- Integrated Microprocessor Control
MCO-5MUV
- Compact Design and Stackable
- Active Background Control with InCu-saFe®
- SafeCell UV Protection
- Direct Heat and Air Jacket System
- Integrated Microprocessor Control
General Specifications
Depth | 21.6 in |
Height | 22.6 in |
Internal Depth | 14.9 in |
Internal Height | 14.8 in |
Width | 18.9 in |
Internal Width | 13.8 in |
Chamber Volume | 49 L |
Weight | 50 lbs |
Shelf Dimensions | 12.2 |
Additional Specifications
ACCESS PORT
Inner Diameter: 30 mm (1.18 inch), on the back side
INSULATION
Extruded Polyethylene
HEATING METHOD
Direct Heat Air (DHA)
AIRFLOW
Gentle Laminar Flow
AIR FILTER
In-line Submicron
UV LAMP
Optional
WATER LEVEL SENSOR
Electronic Reflective (Included)
TEMPERATURE CONTROL RANGE
Ambient temperature +º5C to 50ºC (ambient temperature: 5ºC to 35ºC)
TEMPERATURE UNIFORMITY
±0.25ºC (ambient temperature: 25ºC, setting: 37ºC, CO2: 5 %, no load)
MAXIMUM POWER CONSUMPTION
Max 205 W
TEMPERATURE AND CO2 CONTROL
PID
CO2 CONTROL RANGE
0% to 20%
CO2 CONTROLLER
PID Control System
CO2 GAS CONNECTION
4 mm to 6 mm (0.157 inch to 0.36 inch) diameter tube can be connected
CO2 GAS PRESSURE
0.03 MPaG (0.3kgf/cm2G, 4.3psiG)
CO2 VARIATION
±0.15% (ambient temperature: 25ºC, setting: 37ºC, CO2: 5% no load)
CO2 SENSOR TYPE
TC Sensor
DATA OUTPUT
Optional
DECONTAMINATION METHODS
24 hr full chamber UV decontamination (SafeCell Ultraviolet Light)
CONTAMINATION CONTROLS
Standard 24 hr full chamber UV decontamination (SafeCell Ultraviolet Light) and Copper Alloy Interior Surface (InCu-saFe®)
DISPLAY
Digital Touchpad
HEATING SYSTEM
Direct heat
HUMIDIFYING SYSTEM
Natural evaporation with humidifying pan,
95 %R.H.±5 %R.H.
INTERIOR SURFACES
Stainless steel containing copper
OXYGEN CONTROL RANGE
2% - 18% and 22% - 80%