Features of MCO-170M
- H2O2 Decontamination Technology option safely cleans the incubator chamber in less than 3 hours (Available with MCO-170MUVH-PA)
- Zirconia O2 Sensor exerts precise control on oxygen levels in order to simulate in vivo conditions similar to a cell's native surroundings
- Dual Detector IR2 Sensors offer rapid CO2 recovery without overshoot, even after multiple door openings
- InCu-saFe® copper-enriched stainless steel alloy offers germicidal properties without corrosion, discoloration, or oxidation
- SafeCell UV lamp decontaminates circulating air and the humidity water pan reservoir, with no harm to cultured cells or any ozone production (Available with MCO-170MUV-PA and MCO-170MUVH-PA)
- Direct Heat and Air Jacket System maintains superior temperature uniformity
- Color LCD Touch Panel delivers full control, even with gloved hands
General Specifications
Depth | 28 in |
Height | 35.6 in |
Internal Depth | 20.6 in |
Internal Height | 26.2 in |
Width | 24.4 in |
Internal Width | 19.3 in |
Chamber Volume | 5.7 cu. ft. |
Weight | 170 lbs |
Additional Specifications
ACCESS PORT
1 Access Port on the Rear Upper Left; Diameter = 1.2" (30mm)
INSULATION
Polystyrene
HEATING METHOD
Direct Heat & Air Jacket System
UV LAMP
MCO-170UVS-PA
TEMPERATURE CONTROL RANGE
Ambient +5°C to +50°C, ±0.1°C
TEMPERATURE UNIFORMITY
±0.25°C (Ambient temperature 23°C, setting 37°C, CO2 5%, O2 5%, no load)
CO2 CONTROL RANGE
0% to 20%
CO2 VARIATION
±0.15%
CO2 SENSOR TYPE
Dual Filter IR
CONTAMINATION CONTROLS
SafeCell® UV System and InCu-saFe® Stainless Steel Copper-Enriched Alloy Interior
DISPLAY
LCD Touch Screen
HEATING SYSTEM
Direct Heat & Air Jacket System
INTERIOR SURFACES
Stainless Steel Copper-Enriched Alloy
OXYGEN CONTROL RANGE
1-18% & 22-80%, ±0.2% Stabilized Zirconia Sensor