Beckman Coulter - HIAC PODS
Portable Liquid Particle Counters
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Particle Counters Service ProvidersView All (3)
Documents & Manuals
There are no Documents or Manuals available.
Features of HIAC PODS
General Specifications
| Depth | 17.8 cm |
| Height | 35.6 cm |
| Width | 33 cm |
| Weight | 9.5 kg |
Additional Specifications
Number of Channels
8
Size Channels
ISO-MTD (standard) 4, 4.6, 6, 9.8, 14, 21.2, 38, 68 Μm
ACFTD (optional) ~1, 2, 5, 10, 15, 25, 50, 100 Μm
Flow Rate
50 mL/min standard (consult factory for optional offerings down to 15 mL/min)
Light Source
Laser diode
Calibration
ISO MTD (based on ISO 11171)
Full ISO 11171 or ISO 4402 optional
Counting Efficiency
Meets JIS B9925:1997
Concentration Limit
20,000 particles/mL at 5% coincidence loss (per ISO 11171)
Sample Volume
3 runs (averaged) of 5, 10 or 20 mL (programmable)
Fluid Temp Range
0 to 90°C at 25°C ambient (32 to 194°F at 77°F ambient)
Measured Fluid Temperature
0 to 100°C, ±0.5°C (32 to 212°F, ±0.9°F)
Viscosity Range
10 to 424 cSt
Measurement
10 to 424 cSt ±20% at value
Wetted Materials
Aluminum, stainless steel, sapphire, PTFE and Aflas®
Cleanliness Classification
ISO 4406-1991, ISO 4406.2-1999, NAS 1638, MIL-STD-1246C, NAVAIR 01-1A-1, SAE AS 4059
Data Storage
500 Sample Records
Input/Output
Serial Communication RS-232
Bottle Operation
[i]Purge Volume:[/i] 15 to 30 mL programmable
[i]Cartridge:[/i] CO2, replaceable, rechargeable
[i]Operating Capacity:[/i] 60 samples per cartridge (120 mL sample bottle)
[i]Shop Air:[/i] 60 to 110 psi (4.1 to 7.6 bar) clean, dry
Online Operation
[i]Fluid Pressure:[/i] 40 to 6000 psi (2.75 to 413.7 bar)
[i]Purge Volume:[/i] 15 to 999 mL programmable
Power
DC Input +24 VDC, 2A
AC/Battery Adapter Universal 100 to 240 VAC, 50 to 60 Hz, 60 W
Rechargeable Battery Nickel-Metal Hydride
[i]Operating Time:[/i] 100 samples or 4 hours continuous
[i]Recharge Time:[/i] 2.5 hours
Environment
[i]Ambient Temperature:[/i] 0 to 50°C (32 to 122°F); 20 to 85% relative humidity, non-condensing
[i]Storage:[/i] -40 to 70°C (-40 to 158°F), up to 98% relative humidity, non-condensing