Phase II - Integrated Coating Thickness Gage with Auto Detect
Manufactured by Phase II
The PTG-3500 utilizes an integrated probe that can automatically detect a Ferrous or Non-Ferrous substrate and comes...
The PTG-3500 utilizes an integrated probe that can automatically detect a Ferrous or Non-Ferrous substrate and comes with 2 substrate samples (steel,aluminum), 4 calibrated thickness samples, carry case, batteries and operation manual.
The PHASE II PTG-3500 coating thickness gauge can perform two different methods of calculating thickness measurement by utilizing the characteristics of both eddy current and magnetic induction.
Testing performance is both non-destructive and extremely accurate.
With this state of the art thickness gauge, you can easily detect the thickness of nonmagnetic coating on a magnetic substrate (ferrous) or an insulating coating on a non-magnetic conductive substrate (non-ferrous) utilizing either an integrated probe or our version that comes with 1 auto-detect external probe.
The PHASE II PTG-3500 can be used in many areas of industry including manufacturing, general engineering, commercial inspection, etc.
The PHASE II PTG-3500 coating thickness gauge can perform two different methods of calculating thickness measurement by utilizing the characteristics of both eddy current and magnetic induction.
Testing performance is both non-destructive and extremely accurate.
With this state of the art thickness gauge, you can easily detect the thickness of nonmagnetic coating on a magnetic substrate (ferrous) or an insulating coating on a non-magnetic conductive substrate (non-ferrous) utilizing either an integrated probe or our version that comes with 1 auto-detect external probe.
The PHASE II PTG-3500 can be used in many areas of industry including manufacturing, general engineering, commercial inspection, etc.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Need Equipment Support?
Documents & Manuals
There are no Documents or Manuals available.
Features of Integrated Coating Thickness Gage with Auto Detect
Guaranteed tolerance:
After one-point calibration: +/- 1-3%n or 2µ m (whichever is greater)
Display: 4 digits (digit height = 10mm/0.4″)
Min. measuring area: 0.2″ x 0.2″ (5mm x 5mm)
Min. radius of curvature: Convex: 0.12″ (3mm)
Concave: 1.2″ (30mm)
Min. substrate thickness: Ferrous: 20 mils (0.5mm)
Non-ferrous: 2 mils (50 µ m)
Calibration: Zero Calibration
Foil calibration
Max. Surface temperature of test object: 302 degrees F (contact time max is 2 seconds)
General Specifications
| Depth | 32 mm |
| Height | 161 mm |
| Width | 69 mm |
| Weight | 5 lbs |
Additional Specifications
Measuring range:
0-1000 µ m max. or 0-40 mils
Resolution:
0.1µ m / 0.1mils(0-99µ m) or 1µ m (over 100µ m)
Power source:
4 AA batteries
Videos
Coating thickness gauges Phase II PTG-3500 coating thickness gauges