Sartorius Group - Automatic Mass Comparator with Load Alternator
Mass comparators with automatic load alternator for fully automatic mass comparison of individual reference mass st...
Mass comparators with automatic load alternator for fully automatic mass comparison of individual reference mass standards and groups of weights. These mass metrology devices are suitable for use in precision classes E1 and E2 in national primary laboratories. The advantage of Sartorius mass comparators with automatic load alternator is that special cylindrical or stackable weights are not required.
Sartorius mass comparators with load alternator are suitable for use in precision classes E1 and E2 in national primary laboratories. The comparators perform fully automatic mass comparison of individual reference mass standards and groups of weights. Special cylindrical or stackable weights are not required. Various types of weights and weight groups can be handled universally.
- Fully automatic mass comparator with control unit
- Integrated printer
- Touch screen display
- Filters can be optimally adapted to the particular ambient conditions
- Separate electronic unit to reduce the effect of heat generated during operation
- Backlit graphic display with plain- text displayed for operator guidance and input prompts in English or a choice of 4 other languages
- Linear bar graph showing capacity usage of the electronic weighing range
Active Questions & AnswersAsk a Question
There are no current Discussions
Balances Service ProvidersView All (24)
Documents & ManualsView All Documents
Features
There are no Features available.
General Specifications
Capacity | 51 kg |
Draft Shield | Included |
Linearity | 50 mg |
Readability | 1 mg |
Repeatability | 2 mg |
Additional Specifications
Display: Touch screen display
Data printer: Yes
PC connecting cable: Yes
Control unit: Yes
Test certificates and approvals: Sartorius
Enclosure: Yes
Load alternator: 2
Centermatic: Yes
Application Area: 20 - 50 kg
Electronic weighing range & taring range: 51 kg
Range sensitivity: 5 mg / 50 g
Off-center loading error: 0 mg
Weighing pan dimensions: 70 mm
Sample size (WxD): 120-330 × 300 mm
Typical stabilization time: 10 s
Cycle time ABBA: 720 s
Applications: OIML R111, class E1: 20-50 kg | OIML R111, class E2=== 20-50 kg | OIML R111, class F1=== 20-50 kg | OIML R111, class F2=== 20-50 kg | ASTM E617, class 0=== 20-50 kg | ASTM E617, class 1=== 20-50 kg | ASTM E617, class 2=== 20-50 kg | ASTM E617, class 3=== 20-50 kg
Interface: RS232