Lumex Instruments - CE System CAPEL®-205
Capel®-205 is a new generation capillary electrophoresis system with advanced options and design. Its main features...
Thus Capel®-205 can be the very best choice not only for R&D and low-scale analyses but also for the customers dealing routinely with dozens of samples per day.
High performance capillary electrophoresis is based on the differential migration of components of aqueous samples within a narrow fused silica capillary driven by an electric field. Separated solutes are quantitatively detected at the capillary outlet by a high sensitive optical system based on direct or indirect UV absorbance.
ANALYTICAL CHARACTERISTICS
Detectors
CAPEL®-205 is equipped with a spectrophotometric detector with a wavelength range of 190 - 400 nm.
Light source
Deuterium lamp
Vials
Standard centrifuge-type (e.g.Eppendorf-type)1.5 ml, minimal sample volume 100 µl.
Injection modes
Sample injections by pressure, vacuum or electrokinetically.
Controlled pressure profile of injection with pressure values from – 100 mbar to + 100 mbar.
Electrokinetic injection at (+/-) 1 – 30 kV considering time for voltage ramping.
Programmable injection time / pressure / voltage.
Analysis
With a Voltage from -30 kV to +30 kV with a 1 kV step and an electronic polarity switch.
Current from 0 to 300 µA.
Elforun® software features
Combination of instrument controlling, data acquisition and data processing;
Real time electropherogram visualization;
Electropherogram data processing with overlay mode and basic mathematical operations;
Computation of electrophoresis system parameters;
Customized report output (hard copy and file), data exchanges with worksheets, databases and word processors;
Extensive data processing algorithm customization;
GLP and GALP compliance.
EQUIPMENT AND OPTIONS
Capillary electrophoresis system CAPEL®-205;
Spare capillary cassettes;
Elforun® software package;
Kits for analyses (by request); most of CE-kits of any manufacturers are compatible with Capel®-205.
Active Questions & AnswersAsk a Question
There are no current Discussions
Need Equipment Support?
Documents & ManualsView All Documents
Features of CE System CAPEL®-205
General Specifications
| Depth | 470 cm |
| Height | 530 cm |
| Width | 410 cm |
| Power Requirements | 110/220 Vac, 50/60 Hz |
| Weight | 30 kg |
| Electropheresis Type | Capillary |
Additional Specifications
Detector:
Spectrophotometer
190-400 nm
Analysis:
Constant voltage (+/-) 1 - 30 kV in 1 kV steps
Automatic (electronic) polarity switch
Current 0 - 300 mkA
Pressure (+/-) 1-100 mbar
Programmable changing of wavelength, pressure and voltage during analysis
Injection:
By voltage
By pressure
By vacuum
Sampler:
Autosamplers for 59 vials
Rinsing:
With pressure ~1000 or 2000 mbar
Capillary:
Length: 30-100 cm
Internal diameter: 50, 75, 100 Μm
Capillary thermostating:
Liquid thermostating conducted at temperatures from 10 ºC below to 30ºC above ambient temperature (± 0.1 C)
Controlled parameters:
Wavelength
Pressure
Temperature
Voltage