Lumex Instruments - Liquid Mercury Analyzer RA-915M/RP92

Manufactured by  Lumex Instruments
Follow this Equipment

RA-915M mercury analyzer with RP-92 >attachment is used for determination of mercury concentration in...

RA-915M mercury analyzer with RP-92 attachment is used for determination of mercury concentration in liquid samples (water, blood, urine).

RP-92 attachment is intended to transfer bound Hg from ion to atomic form using “cold vapor” technique.

Measurement procedure

Preliminary mineralization should be carried out by accepted procedure (ISO 5667, 7470, etc.).

Add reduction solution into the bubbler. When integration is on after stabilization of an analytical signal, insert mineralized sample into the reaction vessel. Integration will be finished when the analytical signal goes back to the baseline. Results of analysis of a blank sample should be taken into account.

PRINCIPLE OF OPERATION

Is based on the absorption of the 254-nm resonance radiation by mercury atoms using Zeeman correction for background absorption.  
 
 
 
AREAS OF APPLICATION  
RA-915M mercury analyzer with RP-92 attachment can be used in ecology, sanitary, medicine, and geological-geochemical survey.

Control of mercury content in:

  • waste, natural, and dinking waters;
  • foodstuffs;
  • biological objects (urine, hair, blood, etc.) for diagnosis of risk groups.

Occupational pathology and toxicology surveys:

  • screening population survey;
  • instrumental monitoring of medical treatment of mercury intoxication and further rehabilitation.

Mercury pollution monitoring:

  • monitoring of a demercurization process involving any chemical reagents;
  • quality assurance of the demercurization;
  • determination of mercury content in washouts;
  • determination of mercury content in surface and ground waters.

EQUIPMENT AND OPTIONS

  • RA-915M mercury analyzer;
  • RP-92 attachment;
  • RA-915M software.

SERVICES

On customer’s request our service engineers can carry out installation and commissioning of the Lumex instruments at customer’s site. Consultations and initial training of attendance personnel are provided with due account of customer’s specific needs. Spare parts are delivered and repair is made immediately upon customer’s request (free within the warranty period).

ANALYTICAL CHARACTERISTICS

Subject Detection limit, µg/l Sampling, ml

Water: 0.000520

Blood: 0.51.5

Urine: 0.051

Active Questions & AnswersAsk a Question

There are no current Discussions

Elemental Analyzers Service ProvidersView All (6)

Request Support
Features of Liquid Mercury Analyzer RA-915M/RP92

FEATURES AND BENEFITS   

  • Any sample preparation technique can be used;   * Ability to use single-path or multi-path analytical cell depending on the task (background measurement or analysis of highly contaminated waste water);   * Mercury determination without gold amalgam pre-concentration;   * Wide dynamic measurement range more than 3 exponent (without sample dilution);   * Detection limit is on background level of mercury content in natural waters;   * Visualization of the process of Hg coming from the sample, user-friendly interface;   * Calibration stability (its necessary to check the stability once a day);   * Possibility to carry out field analysis without a PC;   * Measurement of low Hg concentration in drinking and natural water, foodstuffs;    * Control of Hg content in wastewater and washouts;   * Mercury determination in blood, urine and other liquid samples.
General Specifications
Depth 210 mm
Height460 mm
Width 110 mm
Power Requirements 100/240 Vac, 50/60 Hz, 15 W
Weight7.5 kg
Additional Specifications

Size and weight of RP-91: 340x240x120 mm, 4 kg

Power requirements for RP-91: 100/240 Vac, 50/60 Hz, 20 W