Lumex Instruments - Real-Time PCR Analyzer AriaDNA ®

Manufactured by  Lumex Instruments
Follow this Equipment

Polymerase chain reaction (PCR) is now often indispensable technique used in medical and biological research labs fo...

Polymerase chain reaction (PCR) is now often indispensable technique used in medical and biological research labs for a variety of applications including qualitative and quantitative nucleic acid analysis; DNA cloning for sequencing, DNA-based phylogeny, or functional analysis of genes; the diagnosis of hereditary diseases; the identification of genetic fingerprints (used in forensic sciences and paternity testing); and the detection and diagnosis of infectious diseases.

APPLICATION AREAS

  • Clinical and laboratory services for medical, veterinary and phytopathogen diagnostics;
  • Quality control services for food manufacturing, customs control and biological safety;
  • Virus and bacteria diagnostic centers;
  • STD diagnostics.

Microchip-based real-time PCR analyzer for rapid qualitative and quantitative analysis of nucleic acids in various samples.

Microchips for user-designed PCR reagents can be used with various reagents according to the user tasks. PCR mixtures of reagents and samples can be added into microreactors in the same way as in a test tube.

Microchips with lyophilized reagents for the detection and diagnosis of infectious diseases; functional analysis of genes; the diagnosis of hereditary diseases. Microchips with lyophilized PCR reagents can be transported and stored at ambient temperature up to 6 months.

Applications of microchips with lyophilized reagents designed by Lumex:

  • diagnostics of sexually transmitted infections in biological samples (RUO): Trichomonas vaginalis (TV), Candida albicans (Ca), Mycoplasma genitalium (MG), Mycoplasma hominis (MH), Chlamydia trachomatis (CT), Ureaplasma spp (Ur.spp), Neisseria gonorrhoeae (NG) and Herpes simplex virus I/II (HSV);
  • determination of SNP mutations in human genome (RUO) (polymorphisms in genes F2 [20210G/A], F5 [1691G/A], MTHFR [677C/T], [1298A/C], CYP2C9 (*2 and *3), VKORC1 [-1639G/A]);
  • determination of infectious diseases of cows caused by microorganisms (RUO): Chlamydophila pecorum, Chlamydophila abortus, Brucella spp., Ureaplasma diversum, Trichomonas foetus, Campylobacter fetus, Campylobacter jejuni, Listeria monocytogenes, Leptospira interrogans, Mycoplasma bovis, Mycoplasma mycoides;
  • determination of infectious diseases of chicken caused by viruses (RUO): Avian paramyxovirus (NDV), Coronavirus avia (IBV), Birna viridae (IBDV), Reoviridae (AORV);
  • determination of pathogens in raw and processed food (RUO): Salmonella spp. and Listeria monocytogenes.
Active Questions & AnswersAsk a Question

There are no current Discussions

PCR / Thermal Cyclers Service ProvidersView All (14)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of Real-Time PCR Analyzer AriaDNA ®

PCR analysis time (45 cycles) – 20 minutes due to high thermocycling rate (10–12 °C/s);

Low sample and reagent consumption only 0.5–1.8 Μl of 2x PCR master mix for 0.5–1.8 Μl sample;

Low detection limits 1–5 DNA/RNA copies per microreactor;

Microchips with ready-to-use immobilized PCR reagents; reduce user’s labor content;

Qualitative and quantitative DNA/RNA analysis simultaneously in 20–48 microreactors via 2 detector channels (FAM, SYBR Green / ROX);

Minimizing contamination PCR is performed in a microchip isolated from environment;

Real-time data monitoring current process settings (temperature, PCR cycle); analysis time; expected time to complete analysis; PCR curves parameters; plus/minus DNA analysis results;

User friendly graphical interface; Automatic report generation; CFR 21 Part 11 complaint; Export results in industry standard data formats

General Specifications
Depth 300 mm
Height190 mm
Width 250 mm
Power Requirements 90-240 VAC, 50/60 Hz
Weight5 kg
Additional Specifications

Rate of sample thermal cycling  
heating: 12 °C/s 
cooling: 10 °C/s 
 
Minimum DNA content in microreactor: 
15 DNA copies 
 
Sample volume for a single analysis: 
0.5-1.8 Μl 
 
Total PCR analysis time (45 cycles): 
20 minutes 
 
Number of microreactors on a chip: 
20, 24, 30, 48 
 
Reagents required per analysis (2-x PCR mixture): 
0.51.8 Μl 
 
Microchips with immobilized reagents: 
on request 
 
Detection channel 1, dye: 
FAM, SYBR-Green I 
 
Detection channel 2, dye: 
ROX 
 
External PC control: 
USB 2.0 
 
Software designed for PCR analysis, instrument setup and database maintenance: 
Included 
 
Power consumption: 
100 W