TA Instruments - DIL 820 HT

Manufactured by  TA Instruments
Follow this Equipment

High sensitivity optical encoder joined to two high temperature furnace options.

High sensitivity optical encoder joined to two high temperature furnace options. 
 
The 820HT Series operates up to high temperatures in a vertical orientation , making it uniquely set up for the analysis of sintering, studies in Rate Controlled Sintering (RCS) mode and the determination of dilatometric parameters of samples otherwise Impossible to be correctly analyzed on classic dilatometers with horizontal design. 
 
Available in two different models (DIL 821HT and DIL 822HT), both the units feature the new optical encoder with 1nm resolution. That when combined with the True Differential of the DIL822HT, the result is simply the best possible sensitivity and CTE accuracy in the high temperatures segment. 
 
Unlike in horizontal dilatometry the vertical design does not require the sample to be supported. The specimen stands free on the bottom plate and the pushrod is loaded on its tip. 
 
The vertical configuration is particularly well suited for syntherization studies and for all those sample that result difficult if not impossible to be measured with horizontal pushrod dilatometers, like for instance powder samples and samples that during the test develop a vitreous phase. That would lead the sample to interact with the sample holder hence generate friction. A major source of errors in the measurament of dilatometric parameters. 
 
Equipped with a linear motor able to generate a constant force throughout the experiment ,DIL 821HT’s and DIL 822HT’s ensure the contact with sample is maintained linear and with the lightest possible force regardless of the dimensional change.
Active Questions & AnswersAsk a Question

There are no current Discussions

Need Equipment Support?

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features

There are no Features available.

General Specifications

There are no General Specifications available.

Additional Specifications

Sample length 
0 to 25 mm 
 
Sample diameter:  
DIL 821HT: 10mm  
DIL 822HT: 5mm 
 
Material of sample holder:  
Al2O3, Graphite 
 
Contact Force:  
0.01 – 1.0N 
 
Change of length:  
5mm 
 
Resolution AL°C:  
1nm, 0.05°C 
 
Accuracy in Α:  
DIL 821HT: 0.03 x 10-6 K-1  
DIL 822HT: 0.01 x 10-6 K-1 
 
Atmosphere:  
Vacuum, Inert gas, Air (gas unit with massflow controller and vaccum unit included) 
 
Temperature range:  
RT – 2000°C 
300°C – 2300°C

Looking for New or Used Equipment?

Shop on LabX

Shop on Labx.com