Myers-Vacuum - AVC-485
The AVC-485 Automatic Valve Controller is designed to permit either manual or fully automatic operation of diffusion, turbo, or cryopumped high...
The AVC-485 is powered from a 115 volt, 2 amp source and will handle the three basic electropneumatic vacuum valves (roughing, foreline, high vacuum) and a solenoid vent valve in any vacuum system. A screw-driver switch determines the operation mode with separate panel light indicators.
Time delay, for sequential automatic valve operations is preset. Settings can be changed to match a customer’s requirements.
A clear and bold schematic of a typical system is on the front panel to assist in tracing the function being performed. In the automatic mode, LEDs on the schematic indicate valve status.
The panel meter indicates the station pressure in the chamber and at the foreline or regen line. The sensors are the highly reliable and stable GT-034 thermistor tubes.
Operator-Selectable Phases – Pressing the system stop button closes all vacuum valves except the foreline valve on Diffusion/Turbo pump systems, putting the system in a standby condition. Pressing the system vent button puts the chamber in the standby condition and then opens the chamber vent valve. The vent valve is closed automatically after a preset time to conserve backfill gas.
Active Questions & AnswersAsk a Question
There are no current Discussions
Vacuum Gauges / Controllers Service ProvidersView All (5)
Documents & Manuals
There are no Documents or Manuals available.
Features of AVC-485
General Specifications
There are no General Specifications available.
Additional Specifications
Power Input:
115 or 230 VAC, 50/60 Hz
Fuse type and rating:
2 amp 3 AG for 115V service, 1 amp 3 AG for 230V service
Output to energize valves:
same voltage as AC line (valve solenoids must operate from same voltage as AC lines)
Valve solenoid current requirements:
Typical inrush current 0.3 amps
Typical holding current:
0.1 amps
Maximum allowable current:
2 amps
Time delay adjustment range:
all valves 3-20 seconds; auto vent stop 1-15 minutes; external control 5-20 seconds
Range of pressure responsive control circuits:
0-1000 milliTorr
Trip span (differential between turn-on and turn-off):
0-25% of meter scale length
External control relay drive:
5 VDC, 10 milliamps
Pressure sensor tubes (for recorder use):
GT-034
Dimensions front panel:
3-1/2″ high, 19″ wide
Chassis depth:
8″