Myers-Vacuum - GT-034

Manufactured by  Myers-Vacuum
Follow this Equipment

The sensor is encased in a nickel plated, mild steel casing that is baffled so that any vapor from the system does not optically see the element.

The sensor is encased in a nickel plated, mild steel casing that is baffled so that any vapor from the system does not optically see the element.This increases the life of the sensor and has little effect on the amount of time it takes to sense pressure changes. 
 
A zero pot is cleverly hidden in the octal socket to allow individual zeroing of the tube. Each tube is zeroed at the factory and should not require recalibration. This allows for accurate interchangeability between tubes. 
 
The tube can be universally mounted, but is best if mounted pointing down to flow condensable vapors back into the system. 
 
The usable measurement range is from approximate atmosphere to 1 x 10-3 Torr. The casing can easily be pumped to the 10-8 Torr range and up to 60 PSI. The sensor has the most accuracy in the 250 to 1 milli Torr range which is an ideal cross over pressure from mechanical pumps to high vacuum pumps.
Active Questions & AnswersAsk a Question

There are no current Discussions

Vacuum Gauges / Controllers Service ProvidersView All (5)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of GT-034
  • This unique tube is better than standard thermocouple tubes in high vapor and other rugged applications. * Single element utilizes wire that is 10 times thicker than standard thermocouple wire. * Element is glass encapsulated to allow for easy cleaning. * Element is welded to a glass feedthru so there is no epoxy or solder to contaminate your system.
General Specifications

There are no General Specifications available.

Looking for New or Used Equipment?

Shop on LabX

Shop on Labx.com