This product is Discontinued
EPPENDORF - Reference
Manufactured by EPPENDORF
Eppendorf Reference Adjustable-volume Pipette
The patented Reference pipette is synonymous with pipetting accuracy, a smooth stroke, a compact and robust design and superior operating comfort. One multifunctional button operates measuring stroke and blow-out as well as ejects the tip. There are eight adjustable-volume models to cover the range from 0.1 ΜL to 2,500 ΜL, and 31 fixed-volume pipettes are available for the range from 1 ΜL to 2,500 ΜL.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Pipettes Service ProvidersView All (10)
Documents & ManualsView All Documents
Features of Reference
- Fully autoclavable
- One button operation for aspiration, dispensing and tip ejection
- Volume is automatically locked
- TÜV award: user-friendly and user-tested
- Chemically resistant
- Single-handed operation
- Four-digit display
- Three-year warranty
General Specifications
There are no General Specifications available.
Series Modals
EPPENDORF - 0.1-2.5 µL
[+] View Model SpecificationsMaximum Dispensing Volume | 2.5 μL |
Channel(s) | 1 |
Dispensing Volume Range | 0.1 to 2.5 μL |
Increment | 0.002 μL |
Pipette Accuracy | 12.0% 0.25 ?L, 2.5% 1.25 ?L, 1.4% 2.5 ?L |
Pipette Precision | <6.0% 0.25 ?L, <1.5% 1.25 ?L, <0.7% 2.5 ?L |
Pipette Type | Manual Air Displacement |
Pipette Volume Type | Variable |
Additional Model Specifications |
EPPENDORF - 0.5-10 µL
[+] View Model SpecificationsMaximum Dispensing Volume | 10 μL |
Channel(s) | 1 |
Dispensing Volume Range | 0.5 to 10 μL |
Increment | 0.01 μL |
Pipette Accuracy | 2.5% 1 ?L, 1.5% 5 ?L, 1.0% 10 ?L |
Pipette Precision | <1.8% 1 ?L, <0.8% 5 ?L, <0.4% 10 ?L |
Pipette Type | Manual Air Displacement |
Pipette Volume Type | Variable |
Additional Model Specifications |
EPPENDORF - 10-100 µL
[+] View Model SpecificationsMaximum Dispensing Volume | 100 μL |
Channel(s) | 1 |
Dispensing Volume Range | 10 to 100 μL |
Increment | 0.1 μL |
Pipette Accuracy | ±3.0% 10 ?L, ±1.0% 50 ?L, ±0.8% 100 ?L |
Pipette Precision | <0.7% 10 ?L, <0.3% 50 ?L, <0.15% 100 ?L |
Pipette Type | Manual Air Displacement |
Pipette Volume Type | Variable |
Additional Model Specifications |
EPPENDORF - 100-1,000 µL
[+] View Model SpecificationsMaximum Dispensing Volume | 1000 μL |
Channel(s) | 1 |
Dispensing Volume Range | 100 to 1000 μL |
Increment | 1 μL |
Pipette Accuracy | ±3.0% 100 ?L, ±1.0% 500 ?L, ±0.6% 1,000 ?L |
Pipette Precision | <0.3% 100 ?L, <0.2% 500 ?L, <0.2% 1,000 ?L |
Pipette Type | Manual Air Displacement |
Pipette Volume Type | Variable |
Additional Model Specifications |
EPPENDORF - 2-20 µL
[+] View Model SpecificationsMaximum Dispensing Volume | 20 μL |
Channel(s) | 1 |
Dispensing Volume Range | 2 to 20 μL |
Increment | 0.02 μL |
Pipette Accuracy | 3.0% 2 ?L, 1.0% 10 ?L, 0.8% 20 ?L |
Pipette Precision | <2.0% 2 ?L, <0.5% 10 ?L, <0.3% 20 ?L |
Pipette Type | Manual Air Displacement |
Pipette Volume Type | Variable |
Additional Model Specifications |
EPPENDORF - 2-20 µL
[+] View Model SpecificationsMaximum Dispensing Volume | 20 μL |
Channel(s) | 1 |
Dispensing Volume Range | 2 to 20 μL |
Increment | 0.02 μL |
Pipette Accuracy | ±5.0% 2 ?L, ±1.2% 10 ?L, ±1.0% 20 ?L |
Pipette Precision | <1.5% 2 ?L, <0.6% 10 ?L, <0.3% 20 ?L |
Pipette Type | Manual Air Displacement |
Pipette Volume Type | Variable |
Additional Model Specifications |
EPPENDORF - 50-200 µL
[+] View Model SpecificationsMaximum Dispensing Volume | 200 μL |
Dispensing Volume Range | 50 to 200 μL |
Increment | 0.2 μL |
Pipette Accuracy | ±1.0% 50 ?L, ±0.9% 100 ?L, ±0.6% 200 ?L |
Pipette Precision | <0.3% 50 ?L, <0.3% 100 ?L, <0.2% 200 ?L |
Pipette Type | Manual Air Displacement |
Pipette Volume Type | Variable |
Additional Model Specifications |
EPPENDORF - 500-2,500 µL
[+] View Model SpecificationsMaximum Dispensing Volume | 2500 μL |
Channel(s) | 1 |
Dispensing Volume Range | 500 to 2500 μL |
Increment | 2 μL |
Pipette Accuracy | ±1.5% 500 ?L, ±0.8% 1,000 ?L, ±0.6% 2,500 ?L |
Pipette Precision | <0.3% 500 ?L, <0.2% 1,000 ?L, <0.2% 2,500 ?L |
Pipette Type | Manual Air Displacement |
Pipette Volume Type | Variable |
Additional Model Specifications |
Videos