Rudolph Research Analytical - Autopol VII
High Accuracy Polarimeter with Industry Leading Automated Sample Loading.
The Industry Standard Polarimeter
The Autopol® VII combines a pharmaceutical level Polarimeter with fully automated sample loading capability, a perfect answer for high throughput labs that demand accuracy and, reproducibility, combined with versatile sample loading, cell cleaning, rinsing, and drying.
The Autopol® VII has the validation tools, features, and reputation to satisfy today’s global pharmaceutical companies, including: instrument level 21CFR Part 11 compliance, NIST traceable validation and a global list of satisfied customers who praise the Autopol’s® quality, accuracy and reproducibility.
The Autopol® VII offers Automated NIST Traceable 3 point validation combined with fully Automatic TempTrolTM Temperature Control
The Autopol® VII has built-in NIST Traceable Quartz plates to automatically perform a 3 rotation validation on the Autopol VII. To validate the instrument simply press a button to begin the validation routine. Using the on-board quartz plate standards the instrument will perform and record the results of a full 3 point calibration at all wavelengths at the correct temperature.
Active Questions & AnswersAsk a Question
There are no current Discussions
Polarimeters Service ProvidersView All (3)
Documents & Manuals
There are no Documents or Manuals available.
Features of Autopol VII
General Specifications
There are no General Specifications available.
Additional Specifications
Measuring Mode:
Optical Rotation, Specific Rotation, Concentration & User-defined Scales
Measuring Scale:
Degrees Arc, % Concentration
Measuring Range:
±89.9° Arc Optical Rotation, ±999.99° Arc Specific Rotation and 0-99.9% Concentration
Resolution:
0.001° Arc Optical Rotation, 0.001% Concentration, 0.001° Specific Rotation
Reproducibility:
0.002° Arc
Standard Model Accuracy:
0.002° up to 1°, 0.2% up to 5°, 0.01° above 5°
AP Model Accuracy:
(546nm and 589nm): ±0.002° Arc over ±89.9°?Arc. Accuracy for other wavelengths is the same as the standard model.
Prism:
Glan Thompson calcite quartz
Optical Wavelengths:
365nm, 405nm, 436nm, 546nm, 589nm, 633nm (other wavelengths available)
Wavelength Selection:
Automatic by push-button
TempTrol™ Range:
Automatic Electronic Heating & Cooling 15°-35°C
TempTrol™ Accuracy:
±0.2°C
Temp. Probe Range:
10°-40°C
Temp. Probe Accuracy:
±0.1°C
Acid Resistance:
Hastelloy™ measurement cell and Silco Steel™ sample chamber (optional)
Measurement Time:
4°/sec. slew rate and 5 sec. nominal settling time
Light Source:
Tungsten-halogen 6V, 20W, avg. 2,000 hour life
Sample Chamber:
Accepts sample tubes up to 200mm
Data Storage/Internal Memory:
Multimedia Card and/or Secure Digital Card
Communication Interface:
Two RS232 serial ports, one parallel printer port and one auxiliary port
Calibration:
Automatic calibration by push-button
Display:
7.5cm x 10cm graphics LCD, 320 x 240 dots cold fluorescent back lit
User Interface:
Touchscreen
Automatic Sensitivity Control:
Measures samples with transmittance as low as 0.01% (up to O.D. 4.0)
Input Power:
100-240VAC, 50/60 Hz
Operating Dimensions:
35”W x 10.5”H x 17”D 890mm W x 267mm H x 432mm D
Shipping Dimensions:
43”W x 26”H x 23”D 1,025mm W x 625mm H x 625mm D
Operating Weight:
85 lbs. (39kg)
Shipping Weight:
115 lbs. (52kg)