Radionuclides are widely used in many fields of research, nuclear medicine and in life science.
Radionuclides are widely used in many fields of research, nuclear medicine and in life science. Wherever unsealed radioactive materials are used, radiation protection regulations require contamination monitoring for hands, feet and clothing.
The LB 148 is a compact versatile Personal Contamination Monitor for radiation protection applications with back of hand measurement. By means of the new scintillation technology it is possible to measure alpha and beta/gamma contaminations simultaneously but the instrument can also be used as pure alpha or pure beta/gamma monitor. The results can be represented in freely selectable units Bq/cm² or in cps.
The LB 148 has a removable hand probe for clothing measurements. A personal registration can be made with optional card reader or transponder.
Active Questions & AnswersAsk a Question
There are no current Discussions
Elemental Analyzers Service ProvidersView All (6)
Documents & Manuals
There are no Documents or Manuals available.
Features of LB 148 Personal Contamination Monitor
- Low-maintenance ZnS(Ag)-scintillation detectors
- Simultaneous back of hand measurement with automatically moving back of hand detector for optimal measurement geometry
- Space-saving design
- High detector efficiencies
- Removable hand probe for clothing measurement
- Simple operation with graphic display and touch display
- USB, Ethernet, RS 232 and RS 485 interface
- Service functions (plateau, calibration, systematic test,…)
- Permanent data memory for 10000 measuring data
- Extensive nuclide library
- Card reader/transponder for access control (optional)
- Double password hierarchy
- Voice output: Voice instructions, if activated, are output by the monitor
General Specifications
There are no General Specifications available.
Additional Specifications
CPU NXP i.MX 6 ARM Cortex-A9 Windows Embedded Compact 2013
Display Coloured 7" Touch Panel, 800 x 480 pixels
Interfaces 5 USB (Host), 1 USB (Device), 1 Ethernet 10/100/1000Mb, 1 RS232, 1 RS485
External signal transmitters 3 relays max. 50 V, 5 A
Power supply 100 - 240 VAC wide-range input, 50/60 Hz
Power consumption about 14.5 W
Fuse 2 A