Labconco - Protector XL Floor-Mounted
Protector XL Floor-Mounted (Walk-In) Laboratory Hoods have panel liners that may be constructed of a variety of...
They are suitable for general chemistry applications and accommodate oversized apparatus. Widths of up to 16 feet are offered.
Active Questions & AnswersAsk a Question
There are no current Discussions
Fume Hoods Service ProvidersView All (9)
Documents & ManualsView All Documents
Features of Protector XL Floor-Mounted
Features
- By-pass airflow design
- Dry powder, glacier white epoxy-coated steel exterior
- Chemical-resistant molded composite panel liner and adjustable upper and lower baffles with flame spread less than 25 per ASTM E-84
- Removable front and side panels and exterior and interior service access panels for access to plumbing and electrical wiring
- Pre-wired instant-start T8 fluorescent lighting, light and blower switches for 115 volt, 60 Hz operation
- Two 3/16" thick tempered safety glass vertical-rising sashes framed with PVC (4', 5' and 6' models) or epoxy-coated aluminum (8' models)
- Four 3/16" thick tempered safety glass horizontal-sliding sashes with epoxy-coated aluminum frames and sash tracks (8', 10', 12', 16' models)
- Factory prepared for up to eight service fixtures including two gooseneck faucets
- Factory prepared for up to four electrical duplex receptacles
- Shipped in multiple sections, top and bottom
- Epoxy-coated, stainless steel 12.81" ID exhaust connection(s)
- ETL listed
Options
- 4', 5', 6', 8', 10’, 12’, 16’ widths
- Interior depths of 30”, 36” and 48”
Required Accessories
- Remote blowers
- Ductwork
General Specifications
There are no General Specifications available.
Additional Specifications
Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805
Product Subcategory: General Chemistry
Nominal Width: 4'
Electrical Requirements: Domestic
Blower Requirements: Remote blower required
Service Fixtures: 2
Electrical Duplexes: 1
Air Usage/Static Pressure with Sash 50% Open: At 100 fpm, exhausts 680 CFM at 0.16" H2O static pressure
Style: Floor-Mounted
Series Modals
Labconco - 10' Protector XL
[+] View Model SpecificationsAirflow (CFM) | 1360 CFM |
Height | 95 in |
Width | 120 in |
Power Requirements | 115 V, 60 Hz, Domestic |
Lighting | T8 fluorescent |
Face Velocity (FPM) | 100 |
Fume Hood Type | Free-Standing Ductless |
Additional Model Specifications | Product Subcategory: General Chemistry Nominal Width: 8' Blower Requirements: Remote blower required Service Fixtures: 2 Electrical Duplexes: 1 Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805 Air Usage/Static Pressure with Sash 50% Open: At 100 fpm, exhausts 1360 CFM at 0.12" H2O static pressure Depth: 9700901: 39.2" 9700903: 45.2" 9700905: 57.2" |
Labconco - 12' Protector XL
[+] View Model SpecificationsAirflow (CFM) | 1660 CFM |
Height | 95 in |
Width | 144 in |
Power Requirements | 115 V, 60 Hz, Domestic |
Lighting | T8 fluorescent |
Face Velocity (FPM) | 100 |
Fume Hood Type | Free-Standing Ductless |
Additional Model Specifications | Product Subcategory: General Chemistry Nominal Width: 12' Blower Requirements: Remote blower required Service Fixtures: 2 Electrical Duplexes: 1 Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805 Air Usage/Static Pressure with Sash 50% Open: At 100 fpm, exhausts 1660 CFM at 0.15" H2O static pressure Depth: 9701001: 39.2" 9701003: 45.2" 9701005: 57.2" |
Labconco - 16' Protector XL
[+] View Model SpecificationsHeight | 96 in |
Width | 192 in |
Power Requirements | 115 V, 60 Hz, Domestic |
Lighting | T8 fluorescent |
Face Velocity (FPM) | 100 |
Fume Hood Type | Free-Standing Ductless |
Additional Model Specifications | Product Subcategory: General Chemistry Nominal Width: 16' Blower Requirements: Remote blower required Service Fixtures: 2 Electrical Duplexes: 1 Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805 Air Usage/Static Pressure with Sash 50% Open: At 100 fpm, exhausts 2260 CFM at 0.10" H2O static pressure Depth: 9716001: 39.2" 9716003: 45.2" 9716005: 57.2" |
Labconco - 4' Protector XL
[+] View Model SpecificationsAirflow (CFM) | 600 CFM |
Height | 95 in |
Width | 48 in |
Power Requirements | 115 V, 60 Hz, Domestic |
Lighting | T8 fluorescent |
Face Velocity (FPM) | 100 |
Fume Hood Type | Free-Standing Ductless |
Additional Model Specifications | Product Subcategory: General Chemistry Nominal Width: 4' Blower Requirements: Remote blower required Service Fixtures: 2 Electrical Duplexes: 1 Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805 Air Usage/Static Pressure with Sash 50% Open: At 100 fpm, exhausts 680 CFM at 0.16" H2O static pressure Depth: 9860401: 39.2" 9860403: 45.2" 9860405: 57.2" |
Labconco - 5' Protector XL
[+] View Model SpecificationsAirflow (CFM) | 900 CFM |
Height | 95 in |
Width | 60 in |
Power Requirements | 115 V, 60 Hz |
Lighting | T8 fluorescent |
Face Velocity (FPM) | 100 |
Fume Hood Type | Free-Standing Ductless |
Additional Model Specifications | Product Subcategory: General Chemistry Nominal Width: 5' Blower Requirements: Remote blower required Service Fixtures: 2 Electrical Duplexes: 1 Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805 Air Usage/Static Pressure with Sash 50% Open: At 100 fpm, exhausts 900 CFM at 0.24" H2O static pressure Depth: 9860501: 39.2" 9860503: 45.2" 9860505: 57.2" |
Labconco - 6' Protector XL
[+] View Model SpecificationsAirflow (CFM) | 1120 CFM |
Height | 95 in |
Width | 72 in |
Power Requirements | 115 V, 60 Hz, Domestic |
Lighting | T8 fluorescent |
Face Velocity (FPM) | 100 |
Fume Hood Type | Free-Standing Ductless |
Additional Model Specifications | Product Subcategory: General Chemistry Nominal Width: 6' Blower Requirements: Remote blower required Service Fixtures: 2 Electrical Duplexes: 1 Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805 Air Usage/Static Pressure with Sash 50% Open: At 100 fpm, exhausts 1120 CFM at 0.28" H2O static pressure Depth: 9860601: 39.2" 9860603: 45.2" 9860605: 57.2" |
Labconco - 8' Protector with two sashes
[+] View Model SpecificationsAirflow (CFM) | 1560 CFM |
Height | 95 in |
Width | 96 in |
Power Requirements | 115 V, 60 Hz, Domestic |
Lighting | T8 fluorescent |
Face Velocity (FPM) | 100 |
Fume Hood Type | Free-Standing Ductless |
Additional Model Specifications | Product Subcategory: General Chemistry Nominal Width: 8' Blower Requirements: Remote blower required Service Fixtures: 2 Electrical Duplexes: 1 Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805 Air Usage/Static Pressure with Sash 50% Open: At 100 fpm, exhausts 1560 CFM at 0.19" H2O static pressure Depth: 9700807: 39.2" 9700809: 45.2" 9700811: 57.2" |
Labconco - 8' Protector XL with four sashes
[+] View Model SpecificationsAirflow (CFM) | 2200 CFM |
Height | 95 in |
Width | 96 in |
Power Requirements | 115 V, 60 Hz, Domestic |
Lighting | T8 fluorescent |
Face Velocity (FPM) | 100 |
Fume Hood Type | Free-Standing Ductless |
Additional Model Specifications | Product Subcategory: General Chemistry Nominal Width: 8' Blower Requirements: Remote blower required Service Fixtures: 2 Electrical Duplexes: 1 Compliance: SEFA 1-2002, NFPA 45-2000, ASTM E84-01, ASHRAE 110-95, ANSI Z9.5-1993, UL 3101-1/61010-1, CAN/CSA C22.2 No. 1010.1, UL 1805 Air Usage/Static Pressure with Sash 50% Open: At 100 fpm, exhausts 2200 CFM at 0.30" H2O static pressure Depth: 9700801: 39.2" 9700803: 45.2" 9700805: 57.2" |