ELTRA - Carbon / Hydrogen / Sulfur Analyzer CHS-580

Manufactured by  ELTRA
Follow this Equipment

ELTRAs CHS-580 is the ideal analyzer for the simultaneous determination of carbon, hydrogen and sulfur in organic samples.

ELTRA‘s CHS-580 is the ideal analyzer for the simultaneous determination of carbon, hydrogen and sulfur in organic samples.

Thanks to sample weights of 500 mg and more, even inhomogeneous materials can be reliably analyzed. The temperature of the powerful horizontal resistance furnace with ceramic tube can be set in steps from 1 °C to a maximum of 1,550 °C.

The analyzer can be equipped with up to three independent infrared cells according to the user‘s requirements, allowing for a great variety of applications.

Application Examples

Coal, coke, oil, plant materials, rubber, soot, tobacco, waste, ...

Operation CHS-580

After weighing the sample in a ceramic boat, the weight is transferred from the interfaced balance to the PC. If required, sample weights can also be entered manually. Then the ceramic boat is introduced into the furnace for combustion. The average analysis time is 60 to 180 seconds. The detector signals and instrument parameters are displayed during analysis. Evaluation of the signals and display of the results are done automatically; the data can be transferred to a laboratory information management system (LIMS). The CS-580 requires minimum maintenance. The particle filters and chemicals which need to be maintained are easily accessible.

Measuring Principle CHS-580

In the CHS-580 the sample is burnt in an oxygen atmosphere at temperatures up to 1,550 °C. The furnace temperature can be freely selected in steps of 1 °C. The combustion gasses (CO2, H2O, SO2) coming from the furnace and first pass through a dust filter and then into the heated H2O infrared cell. After the water vapor is chemically absorbed, the dried CO2 and SO2 gas is detected in the additional infrared cells. Depending on the configuration, it is possible to combine up to three infrared cells with different sensitivities.

Active Questions & AnswersAsk a Question

There are no current Discussions

Gas and Air Analyzers Service ProvidersView All (10)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of Carbon / Hydrogen / Sulfur Analyzer CHS-580
  • simultaneous carbon, sulfur and hydrogen determination with minimal sample preparation
  • wide range of organic materials can be analyzed
  • rapid, precise, accurate and reliable element determination
  • resistance furnace can be set up to 1550°C in steps of 1°C
  • customized infrared cells provide wide, dynamic measuring range
  • due to gold IR path, increased cell live time for analysis of halogen or acid containing samples
  • powerful software (multilingual, customized display, export of results)
  • single and multipoint calibration
  • low maintenance
  • robust design allows usage in production control and laboratory
General Specifications

There are no General Specifications available.

Additional Specifications

  • Measured elements carbon, hydrogen, sulfur
  • Samples organic
  • Furnace alignment horizontal
    Sample carrier ceramic boats
  • Field of application agriculture, biology, chemistry / plastics, coal / power plant, construction materials, environment / recycling, medicine / pharmaceuticals
  • Furnace resistance furnace (ceramic tube), adjustable up to 1550°C (steps of 1°C)
  • Detection method solid state infrared absorption
  • Number of IR cells 1 - 3
  • Material of IR path gold
  • Typical analysis time 60 - 180 s
  • Chemicals required magnesium perchlorate, sodium hydroxide
  • Gas required oxygen 99.5% pure (2 - 4 bar / 30 - 60 psi)
  • Power requirements 230 V, 50/60 Hz
    max. heat up current 2000 W
  • Dimensions (W x H x D) 55 x 80 x 60 cm
  • Weight ~70 kg
  • Required equipment balance (resolution 0.0001g), monitor, PC
  • Optional accessories TIC module, voltage stabilizer 5 KVA

Looking for New or Used Equipment?

Shop on LabX

Shop on Labx.com