Carbolite - Split tube furnace up to 1700°C - HTRV-A
The HTRV-A split tube furnaces have a maximum operating temperature of 1700 °C.
The split heating module allows either easy positioning of the work tube or positioning around reactors which have fixed end flanges. The split design may also allow faster cooling of samples. The control thermocouple is fitted in the centre of the heating zone. Cooling channels are engineered into the housing to aid with convection cooling of the outer case.
The two furnace chamber halves consist of high grade insulation plates with vertically hanging MoSi2 heating elements. A safety switch protects the operator by switching off the heating elements once the furnace is opened.
The tube furnace is supplied without a stand, allowing customers to build them into their own equipment. Optional ‘L’ stands are available allowing the furnaces to be self supporting.
Application Examples
annealing, carbonisation, crystal growth, debinding, degassing, drying, glowing, hardening, metal injection moulding (MIM), pyrolyses, rapid prototyping, sintering, sublimation, synthesis, tempering
Active Questions & AnswersAsk a Question
There are no current Discussions
Lab Furnaces Service ProvidersView All (8)
Documents & Manuals
There are no Documents or Manuals available.
Features of Split tube furnace up to 1700°C - HTRV-A
General Specifications
There are no General Specifications available.
Additional Specifications
HTRV-A __/70/120
Max temp (°C) 1600
Max outer diameter accessory tube (mm) 70
Heated length (mm) 120
Dimensions:
External H x W x D (mm) 700 x 700 x 890*
Furnace weight (kg) 65
Tube length for use in air (mm) 470
Tube length for use with modified atmosphere (mm) 910
Control module dimensions H x W x D (mm) 850 x 560 x 500
Control module weight (kg) 60
Uniform length ±5°C (mm) 50
Power (W) 4800
HTRV-A __/70/250
Max temp (°C) 1600,1700
Max outer diameter accessory tube (mm) 70
Heated length (mm) 250
Dimensions:
External H x W x D (mm) 800 x 600 x 890*
Furnace weight (kg) 75
Tube length for use in air (mm) 600
Tube length for use with modified atmosphere (mm) 1040
Control module dimensions H x W x D (mm) 850 x 560 x 500
Control module weight (kg) 90
Uniform length ±5°C (mm) 125
Power (W) 6000
HTRV-A __/70/500
Max temp (°C) 1600,1700
Max outer diameter accessory tube (mm) 70
Heated length (mm) 500
Dimensions:
External H x W x D (mm) 1050 x 700 x 890*
Furnace weight (kg) 120
Tube length for use in air (mm) 850
Tube length for use with modified atmosphere (mm) 1290
Control module dimensions H x W x D (mm) 850 x 560 x 500
Control module weight (kg) 90
Uniform length ±5°C (mm) 250
Power (W) 8000