Analytik Jena - SPEKOL 1300/1500
Manufactured by Analytik Jena
The little giant with great performance
The UV VIS spectrophotometers SPEKOL® 1300/1500 are single beam systems for the range of 190-1100 nm with an outstanding price-performance ratio. Robust in design, the instruments are compact, light and easy to use. A constant spectral bandwidth of 4 nm is maintained over the entire wavelength range.
Active Questions & AnswersAsk a Question
1Replies
DNA quatification
1Replies
Light source errors
UV/VIS Spectrophotometers Service ProvidersView All (32)
Documents & ManualsView All Documents
Features of SPEKOL 1300/1500
- High-resolution LCD-VGA screen providing superb images and easy operation
- Water-proof touch-sensitive keyboard
- Numerical backlit display (SPEKOL® 1300 ) or graphical backlit display (SPEKOL® 1500)
Internal software:
- Measurement of transmittance, absorbance or concentration via factor, simple standard or linear calibration
- Kinetic measurement
- Printing and storage of methods and measurement results
- Determination of purity and concentration of DNA and protein at 260 and 280 nm
General Specifications
There are no General Specifications available.
Series Modals
Analytik Jena - SPEKOL 1300
[+] View Model SpecificationsBandwidth | 4 mm |
Depth | 365 mm |
Height | 175 mm |
Width | 465 mm |
Power Requirements | adjustable 115 / 230 VAC, ±10% 60/50 Hz |
Photometric Range | 0-125.0 % T, 0-3.000 A, 0-9999 C |
Weight | 11.5 kg |
Wavelength Range | 190 to 1100 nm |
Scanning Speed | Fast, medium, slow |
Optics | Single Beam |
Photometric Noise | ± 0.002 A |
Stray Light | </= 0.3 % T at 220 nm and 340 nm |
Additional Model Specifications | Optical Performance Wavelength Accuracy: ± 2 nm Repeatability: ± 1 nm Spectral slit width: 4 nm Photometric Performance Photometric Mode: Transmittance, Absorbance, Concentration Accuracy: ± 0.004 A at 0.5 A Repeatability: ± 0.003 A at 0.5 A Baseline: ± 0.004 A at 500 nm Drift: ± 0.002 A/h (at 500 nm after 1 hour warm up) |
Analytik Jena - SPEKOL 1500
[+] View Model SpecificationsDepth | 365 mm |
Height | 175 mm |
Width | 465 mm |
Power Requirements | 115 V / 230 V (adjustable), 50 to 60Hz |
Photometric Range | 0 %T to 125.0 %T, -0,1 A to 3 A, 0°C to 9999°C |
Weight | 11.5 kg |
Wavelength Range | 190 to 1100 nm |
Optics | Single Beam |
Stray Light | </= 0.3 %T at 220 nm and 340nm |
Additional Model Specifications | Optical Performance Wavelength Accuracy: ± 2 nm Repeatability: ± 1 nm Spectral slit width: 2 nm Photometric Performance Photometric Mode: Transmittance, Absorbance, Concentration Accuracy: ± 0.01 A Repeatability: ± 0.003 A at 0.5 A Baseline: ± 0.004 A at 500 nm Drift: ± 0.002 A/hr (at 500 nm after 1 hour warm up) |