Tunability Equals Flexibility      
      
No filters are needed for the VersaMax™ microplate reader. The reader uses a grating monochromator to select the exact wavelength you need for every assay. With a wavelength range from 340 to 850 nm, it's like having 510 filters free.      
      
Exceptional Performance      
      
The VersaMax reader uses the best plate for your assay. The narrowly focused light beam and optical design give you the same high performance with round bottom, flat bottom, or half area well plates. The dual-wavelength readings report the actual absorbance at each wavelength, so you can analyze the data the way you want. The reader also provides controlled temperatures, up to 45 °C, for kinetics at ambient and elevated temperatures.      
      
Affordable      
      
With the VersaMax microplate reader, you will never have to buy another filter. SoftMax® Pro, the industry-leading microplate reader software, is included at no additional charge. And the VERSAmax reader is robotics-compatible.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Asked bySafia

avatar placemark

Updated bySoftMaxMan

Microplate Readers Service ProvidersView All (12)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of VersaMax
  • Wavelength range from 340 to 850 nm      
  • Dual-wavelength readings report actual absorbance at each wavelength      
  • Controlled temperatures up to 45°C for kinetics at ambient and elevated temperatures      
  • Same high performance for round bottom, flat bottom or half area well plates      
  • SoftMax® Pro industry-leading microplate reader software included      
  • Robotics-compatible      

Applications      

  • Microbial growth/MIC      
  • IC 50s/LD 50s      
  • Endpoint ELISAs/EIAs      
  • Cytoproliferation/cytotoxicity      
  • Colorimetric protein      
  • Kinetic ELISAs/enzyme assays      
  • Platelet aggregation      
  • Bacterial identification      
  • Clotting/clot lysis
General Specifications
Bandwidth2 nm
Depth 38 cm
Height22 cm
Detection ModeAbsorbance
Width 58 cm
Photometric Range0.000 to 4.000 OD
Weight30 lbs
Wavelength Range340 to 850 nm
Wavelength SelectionMonochromator, tunable in 1-nm increments
Microplate Size96-well
SoftwareExternal
Temperature Control 1
Additional Specifications

Photometric linearity (405 nm): 0 to 3.000 OD      
Photometric accuracy (340-850 nm): < ±0.006 OD ±1.0%, 0-2.0 OD      
Wavelength accuracy: < ±0.003 OD ±1.0%, 0-2.0 OD      
Endpoint read time: 12 seconds      
Kinetic read time: 9-second minimum interval      
Temperature uniformity (microplate): ±0.5 °C @ 37 °C well-to-well