AMETEK - Jerome 451
Gold Film Fence-line Mercury Monitoring System
Unattended
The 451 is designed for set-and-forget mercury vapor monitoring, inside or out.
Flexible
A detachable hand-held mercury vapor analyzer is contained within the unit for pinpointing a gas source.
Smart Alarms
Define alarm points that meet your needs and receive email notifications and warnings.
Wireless
Measurement data is transmitted wirelessly from 451 stations to the user’s PC.
Rugged
The 451 is housed in a steel casing, insulated, air conditioned and ready for nearly any environment.Unattended
The 451 is designed for set-and-forget mercury vapor monitoring, inside or out.
Flexible
A detachable hand-held mercury vapor analyzer is contained within the unit for pinpointing a gas source.
Smart Alarms
Define alarm points that meet your needs and receive email notifications and warnings.
Wireless
Measurement data is transmitted wirelessly from 451 stations to the user’s PC.
Rugged
The 451 is housed in a steel casing, insulated, air conditioned and ready for nearly any environment.
Active Questions & AnswersAsk a Question
There are no current Discussions
Need Equipment Support?
Documents & ManualsView All Documents
Features of Jerome 451
General Specifications
There are no General Specifications available.
Additional Specifications
Detection Range: 0.003 mg/m3 to 0.999 mg/m3
Resolution: 0.001 mg/m3
Accuracy: ±8% at 0.100 mg/m3
Sample Intervals: 1, 2, 5, 10, 15, 30, 60 or 120 minutes
Regeneration Intervals: 6, 12, 24 and 48 hours
Flow Rate: 750 mL/min
Operating Environment: -40 to +55°C, non-explosive, 0-100% relative humidity
Result Units: mg/m3
Data Storage Capacity: 50,000 readings
Power Requirements (451 Unit): 100-120 VAC, 50/60 Hz, 7 A or 210-240 VAC, 50/60 Hz, 4 A
Control Board Fuse: Fast-acting IEC 1.6 A 250 V, 5 mm x 20 mm
Case Construction: Polyester with insulation to reduce heat transfer
Battery (Detached 431-X Unit): Rechargeable NiCad 6 hour life, charges in 14 hours *only applicable when 431-X unit is not attached to 451
Estimated Sensor Life: 3 to 6 months depending on operating conditions
Display: 8 digit alphanumeric LCD
Output: Spectra Radio
Analog: 4-20 mA current loop
Dimensions: 26″ W x 25″ H x 9″ D (67 cm W x 64 cm H x 23 cm D)
Weight: 52 lbs. (24 kg)
Warranty: 1 year, factory parts and labor except for 431-X sensor