AMETEK - Jerome J605
Gold Film Hydrogen Sulfide Analyzer
The J605 is housed in a light and ergonomically designed case, setting a new precedent for low-level, portable monitors. The industry-proven, gold film sensor has been upgraded and improved from previous models, resulting in longer life expectancies.
Ultra Sensitive
Detect hydrogen sulfide in air at levels as low as 3 parts per billion (3ppb).
Proven
The Jerome’s gold film sensing technology has been validated in the laboratory and in the field for more than 35 years.
Rugged
Sturdy metal housing and solid-state design make the J605 tough enough for nearly any environment.
Reliable
With 18 hours of battery life per charge, the J605 is always ready when you need it.
Smart
Features on-board data logging for up to 20,000 measurements, USB data transfer, and SCADA interface capabilities through 4-20ma output.
Active Questions & AnswersAsk a Question
There are no current Discussions
Need Equipment Support?
Documents & ManualsView All Documents
Features of Jerome J605
General Specifications
There are no General Specifications available.
Additional Specifications
Detection Range: 3 ppb (0.003 ppm) – 10 ppm
Resolution: 20 ppt (0.02 ppb) *range dependent
Accuracy: ±1 ppb at 5 ppb ±3 ppb at 50 ppb ±0.03 ppm at 0.5 ppm ±0.3 ppm at 5.0 ppm
Response Time: Survey Mode: 12-27 seconds
Manual Mode: 12-27 seconds
Auto Mode: 12-53 seconds
Flow Rate: 150 mL/min
Result Units: ppm, ppb
Integrated Data Storage: 20,000 samples; retains date, time and location
Display: 3.5” (9 cm) backlit LCD display
Battery: Rechargeable NiMH 18 hour life, charges in 3 hours
Power Requirements: 12 VDC for the instrument 100-240 VAC, 47-63 Hz, 3.2 A for the AC power supply/charger
Fuse: Auto-resetting fuse
Operating Environment: 0°C to 40°C, non-condensing, non-explosive
Output: Digital: USB serial data to PC, printer or USB flash drive
Analog: 4-20 mA passive current loop; accurate to 0.3% of output
Dimensions: 11″ L x 6″ W x 6.5″ H (28 cm L x 16 cm W x 17 cm H)
Weight: 5.4 lbs. (2.5 kg)
Certifications: TUV 61010, CE