Waters - 432 Conductivity Detector

Manufactured by  Waters
Follow this Equipment

Features a multi-electrode flow cell that offers exceptional sensitivity and stability for single-column or suppressor-based ion chromatography.

The 432 Conductivity Detector provides solutions for chemically-suppressed or single-column ion chromatography. Use it as a stand-alone module or configure it into an HPLC system that links to chromatography data workstations:

The 432 Detector’s patented five-electrode flow cell design:

  1. Reduces noise
  2. Enhances sensitivity
  3. Allows for direct and indirect conductivity detection independent of eluent background conductivity
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Asked byMothusi

avatar placemark

Updated bylae

avatar placemark

Asked byAng

Conductivity Detector Service ProvidersView All (21)

Request Support
Features of 432 Conductivity Detector
  • High sensitivity and stability due to its multi-electrode flow cell and user-programmable temperature control, which automatically adjusts the oven temperature to minimize baseline drift
  • Reduced flow path volume for narrow-bore applications
  • Increased operational uptime with self-diagnostics
  • Easy, unattended routine analyses with autozero and autoranging capabilities
  • Sophisticated circuitry efficiently isolates detection electrodes from sources of transient electronic noise for optimal sensitivity
  • Autozero feature automatically sets the baseline to zero at the start of each run eliminating the need for manual adjustments
  • Autorange ensures a precise conductivity base range
  • Leak sensors alerts user if leak is detected in flow cell

Low noise equals high sensitivity

The low noise performance of the 432 detector ensures that you obtain an exceptional range of sensitivity. The sophisticated circuitry of the 432 detector efficiently isolates detection electrodes from sources of transient electronic noise for optimal sensitivity.

Noise specifications often vary among other conductivity detectors. When noise specifications are based upon 18 MO water (conductivity of approximately 1-2 µS), the result is artificially low noise. During real analyses, the background conductivity is somewhat greater (chemically suppressed, 10-16 µS) or significantly greater (single column, 200-2000 µS) than zero. Therefore, the noise specification measured at low conductivity is often misleading. The 432 noise specification of less than 0.005 µS/cm is measured using 1mM KCl (1 mL/min.) with a background conductivity of 147 µS, assuring ultimate sensitivity with Waters 432 Conductivity Detector whether you use single-column or suppressor-based ion chromatography.

General Specifications

There are no General Specifications available.

Additional Specifications

Drift: Less than 0.05 µS/hr/°C (Ambient)
(Base 200 µS, Sensitivity: 0.005, Eluent: 1 mM kCl)
Noise: Less than 0.005 µS/cm
(Base 200 µS, Sensitivity: 0.005, Eluent: 1 mM kCl)
Temperature control: Front-panel selectable: OFF, 30, 35, 40, 45, 50, 55, 60, 65°C
Temperature performance: 0.5°C/hr
Response time: Fast: 0.25 sec, Standard: 0.5 sec, Slow: 1.0 sec
Cell volume: 0.6 µL
Wetted materials: 316 stainless steel, PTFE, and PCTFE
Operating pressure: 70 psi maximum
Height: 9.4 inches
Length: 21 inches
Width: 11.5 inches
Weight: 17.7 lbs
Operating temperature: 4 to 35°C
Operating humidity: 20 to 80% RH, noncondensing
Voltage range:
115 Vac: +/- 10 Vac
230 Vac: +/- 10 Vac
Line frequency ranges:
50 Hz: +/- 2 Hz
60 Hz: +/0 2 Hz
120 Vac fuse rating: 2 A
240 Vac fuse rating: 1 A
Recorder output: 0 to 10 mV
Integrator output: 10, 50, 100 µS/1V FS
Marker output: Isolated contact output
Controller bus: IEEE-488, Powerline
Method: Five pole system
Conductivity range: 0-10,000 µS/cm
Autozero: Manual (front panel), Automated (rear panel contact closure)
Autobase: Manual (front panel), Automatic (Powerline control)
Polarity: Front panel indicator light
Leak sensor: Front panel warning and audible signal
All performance specifications are for the 432 detector following a minimum warm-up period of three hours.