Agilent Technologies - 5975C Series GC/MSD
Performance, productivity and confidence.
Active Questions & AnswersAsk a Question
GC-Single Quad/MSD Service ProvidersView All (1)
Documents & ManualsView All Documents
Features of 5975C Series GC/MSD
- Inert ion source—now programmable up to 350°C—delivers enhanced response for active compounds and late eluters
- Proprietary hyperbolic quadrupole enhances performance up to 1050 u; the unique stability of the quartz monolith allows the analyzer to be heated to 200°C to eliminate contamination commonly seen at lower temperatures
- Triple-Axis HED-EM Detector provides the market’s highest signal-to-noise specification (1 pg OFN, 400:1)
- Trace Ion Detection and the second generation of Deconvolution Reporting Software (DRS) accurately identify and quantify trace components that coelute with other analytes or matrix peaks
- Synchronous SIM/Scan mode lets you selectively monitor for ions of interest with high sensitivity SIM while simultaneously acquiring library-searchable scan data
- Safe operation with low cost, hydrogen carrier gas—with sensitivity greater than most other GC/MS systems
- All ionization modes (PCI, NCI and EI) in one automated sequence using the standard Cl source; AutoCI feature makes CI as easy as EI
- High performance and entry level models that are perfectly complemented by the Agilent 7890A GC, 6890 GC, and 6850 GC
- MSD Productivity ChemStation for most laboratories and MSD Security ChemStation for laboratories adhering to FDA’s 21 CFR Part 11 guidelines
- Control of two GC/MS systems from a single MSD ChemStation
- Integrated control of the versatile CTC PAL autosamplers can boost your lab’s output with automated sample preparation (available option).
- Retention Time Locking (RTL) gives you repeatable retention times across systems regardless of operator, detector type and column maintenance. In combination with Agilent or user-created databases, it will let you quickly screen unknown peaks with added confidence.
- Automatic tuning to ensure more accurate results every time for EI, PCI and NCI Modes.
- SemiQuant capability to estimate the concentration of uncalibrated compounds.
- Integration with Enterprise Content Management (ECM) makes it easy to store, retrieve and organize all of the electronic data generated in your laboratory – including your GC/MSD files and reports.
- 5975C VL MSD — Economical, small-footprint system for routine analyses
- 5975C inert MSD — High performance, full-featured system for demanding analyses
- 5975C inert XL MSD — High performance, full-featured system optimized for complex EI analyses
- 5975C inert XL EI/CI MSD — Highest level of GC/MS performance and productivity available
General Specifications
| Depth | 56 cm |
| Detector | Triple-Acis HED-EM with extended life EM |
| Height | 50 cm |
| Width | 88 cm |
| Scan Speed | 12500 amu/sec |
| Weight | 96 kg |
| Mass Spectrometer Type | Quadrupole Mass Analyzer |
Additional Specifications
[u]Mass Selective Detector[/u]
Mode (standard): EI
Modes (optional): PCI, NCI, EI with CI source
Ion source type: Noncoated inert EI source for turbomolecular pump systems (optional for diffusion pump system). Stainless steel EI source for diffusion pump system.
Mass filter: Monolithic hyperbolic quadrupole
Maximum mass: 1050 u
Pumping system: 65 L/s diffusion pump, 70 L/s or 262 L/s turbomolecular pump with 2.5 m3/h mechanical pump
[u]Installation Checkout Specifications[/u]
EI scan sensitivity: 1 µL injection of a 1-pg/µL OFN standard scanning from 50 to 300 u will give a nominal m/z 272 ion
With helium: S/N 400:1 for turbo systems, 200:1 for the diffusion system
With hydrogen: S/N 100:1 for turbo systems and 50:1 for the diffusion system
PCI scan sensitivity: 1 µL injection of a 100 pg/µL BZP standard scanning from 80 to 230 u will give 125:1 S/N with methane and 500:1 with ammonia at nominal m/z 183 ion
NCI scan sensitivity: 2 µL injection of a 100 fg/µL OFN standard scanning from 50 to 300 u will give 600:1 S/N with methane and 300:1 with ammonia at nominal m/z 272 ion.
[i]See PDF for further specifications.[/i]
Videos
7890A GC and 5975C