Features of T50
- Terminal with brilliant Touch Screen
- Expandable with dispensing burette drives
- Comprehensive communication connections
- Preconfigured system application packages
General Specifications
There are no General Specifications available.
Additional Specifications
Potentiometric sensor inputs
Measurement range: ± 2000 mV
Error limit: 0.2 mV
Polarized Sensor input
Measurement range: 0 … 2000 mV / 0 … 200 µA
Resolution: 0.1 mV / 0.1 µA
Error limit: 0.2 mV / 0.3 µA
Temperature sensor input PT1000
Resolution: 0.1 °C
Error limit: 0.2 °C
Burette drive
Burette resolution (for 10-mL burette): 0.5 µL (1/20’000 of the burette volume)
Error limits: At 10 % 15 µm, at 30 % 15 µm, at 50 % 25 µm, at 100 % 50 µm
Resolver: 32 pulses/mm stroke
Resolver resolution: 0.0625 % of burette volume
Filling time and discharge time: 20 s at 100 % filling rate
Terminal
Control: Touch control
Screen: 1/4 VGA 5.7” color TFT
Resolution: 320 x 240 pixel
Angle: Adjustable in three positions
Series Modals
METTLER TOLEDO - T50 Titrator
[+] View Model SpecificationsTitrator Type | Potentiometric +KF |
Additional Model Specifications | Automatic burette recognition: yes Short Cuts: yes, max. 12 per user Number of burette drives: 1 internal +3 external (optional, full titration control) Upgradable: to T90 Flexible method creation: extended flexibility KF Titration: yes Sensor boards: max. 2 Series sequences: no Task list (number of tasks): max. 10, max. 7 parallel Hot Plug and Play: yes Plug&Play Sensors: yes |