METTLER TOLEDO - InLab Reach
Manufactured by METTLER TOLEDO
A glass-body, combination pH electrode with S7 screw head and 300 mm shaft. Designed for accurate measurement in a wide range of sample types.
The electrolyte in conventional silver/silver chloride reference systems must be saturated with silver chloride.
With critical media such as those containing sulphides or proteins, this will result in diaphragm contamination with silver sulphide. This can not happen with InLab® electrodes which have an ARGENTHAL™ lead off and silver ion trap that ensure the electrolyte remains completely free from silver ions. The ARGENTHAL™ reference system also improves service life and measurement accuracy at differing sample temperatures.
The glass shaft provides resistance against organic and corrosive samples. Different cables can be connected to the S7 screw head allowing use with a wide range of meters.
With critical media such as those containing sulphides or proteins, this will result in diaphragm contamination with silver sulphide. This can not happen with InLab® electrodes which have an ARGENTHAL™ lead off and silver ion trap that ensure the electrolyte remains completely free from silver ions. The ARGENTHAL™ reference system also improves service life and measurement accuracy at differing sample temperatures.
The glass shaft provides resistance against organic and corrosive samples. Different cables can be connected to the S7 screw head allowing use with a wide range of meters.
Active Questions & AnswersAsk a Question
There are no current Discussions
pH Meters Service ProvidersView All (13)
Documents & ManualsView All Documents
Features of InLab Reach
- ARGENTHAL™ reference
- Silver ion trap
- Glass shaft
- Detachable cable
- 300 mm shaft
General Specifications
| pH Range | 0 to 14 |
Additional Specifications
Reference system: ARGENTHAL™ with Ag+ -trap
Shaft length: 300 mm
Shaft material: Glass
Temperature range °C: 0 ... 100
Cable and connections: S7
Type of membrane glass: HA
Shaft diameter: 12 mm
Storage longterm: 3 mol/L KCl (52340049) in watering cap
Reference electrolyte: 3 mol/L KCl
Storage shortterm: 3 mol/L KCl, buffer pH 4 or 7
Response time (98% between pH 4...7): < 20 sec.
Membrane resistance (25 °C): < 600 MOhm