INFORS HT - Multitron 2
Space in the laboratory is always scarce and expensive. The Multitron 2 combines flexibility and working reliability...
Active Questions & AnswersAsk a Question
Shakers Service ProvidersView All (14)
Documents & ManualsView All Documents
Features of Multitron 2
General:
The Multitron can be adapted to changing needs to provide the optimal solution regarding capacity and incubator operation.
The tray is unlocked automatically after stopping and can then be moved onto the open door by pulling forwards and thereby offers free access to the cultures. This process is aided by roller bearing on the door for easy slding of the tray in and out. So, one does not have to work in the incubation chamber. High-grade steel ball bearings facilitates easy moving of the table. Rapid removal of flasks is possible also without pulling the table out. Upon closing the door, the tray is automatically locked.
Penta-Drive: The shaking table is secured by five points of attachment for a smooth, reliable shaking action. The unique combination of one fixed and four elastic mountings prevent alignment problems and the generation of noise which are inevitable in the usual triple-excentric designs.
Rapid removal of flasks is possible also without pulling the table out.
The lowest unit (or an individual unit, which is installed under a laboratory bench), can likewise be comfortably loaded and unloaded by withdrawing the table.
In contrast to chest type free-standing machines, the top surface can be used as a working area or even for locating light laboratory instruments.
Precise temperature control; option to use a mobile temperature sensor for measurement directly in liquid.
The Multitron offers the largest shaking capacity of its class: The type M table (85 x 47 cm) has room for e.g. 87 x 100 ml or 6 x 5000 ml Erlenmeyer flasks.
The highest table simply rests on the opened door with a user-friendly working height of 1,30 m . Individual flasks can be handled comfortably. With heavily loaded tables, this is far less troublesome for the user than a free-standing system. Also, with the lower unit, the open door serves as a shelf and so safeguards the backs of the users.
The design gives the greatest possible consideration to floor loading, installation and usability (e.g. for a 3-stack unit. the height is under 2 meters and it weighs less than 450 kg).
Also, stacked models all use the same drive construction, which guarantees accurate and vibration-free running despite the high forces involved.
Safety features:
Shaking speed over-range cut out
Adjustable over-temp setting
Self -monitoring with error messages
All electrical components segregated from the incubation chamber and possible spills
Auto-stop of shaking on door-opening
Door interlock to restrict access until the shaking tray stops
Auto-restart in the event of power failure
Indication of power failure
Slow start and stop to prevent spill over of cultures
Alarm output contactors provided as standard for remote monitoring
Option of remote monitoring of key values
Designed for working heights in stacked systems to minimise bending or stretching of the user
Downwards-opening door allows handling of heavy trays without lifting
General Specifications
Depth | 860 mm |
Height | 550 mm |
Maximum Shaking Speed | 400 rpm |
Width | 1070 mm |
Maximum Temperature | 65 °C |
Power Requirements | 115/230V ± 10%, 50/60Hz |
Temperature Resolution | 0.1 °C at the pt-100 |
Weight | 90 kg without cooling |
Shaking Speed | 30–400 RPM |
Temperature Accuracy | 0.2 °C in liquid |
Additional Specifications
Throw: ø 25 mm or 50 mm
Accuracy: 0,1% of maximum speed
Interface: RS232 bi directional