Features of UltiMate™ 3000 Basic Automated System
Recommended Use For moderate throughput laboratories that need LC and entry level UHPLC capabilities and are investing in their first systems. As an LC/UHPLC starter kit this automated modular system can be extended with additional detectors, a more spacious column compartment, or even access to column switching techniques.
Typical application areas are teaching laboratories or quality control laboratories with moderate sample throughput requirements. The quaternary pump is recommended for methods deploying either conventional 3-5µm, or UHPLC sub-2µm particle columns where a moderately fast gradient response is acceptable for the benefit of increased mobile phase flexibility (i.e., multiple solvents, extended column wash, ternary or quaternary gradients). A four-channel solvent degasser is integrated in the pump module to simplify operation.
Important Capabilities/Features:
Up to 10 mL/min flow rate supporting up to 62 MPa (9000 psi/620 bar) backpressure for column format and column particle size flexibility * Fast and stable column thermostatting between 5°C above ambient and 50°C improves retention time precision. Compact and integrated autosampler column compartment to automate injections of multiple samples and calibration standards * Obtain near-zero dead volume for optimal peak shapes even over extended periods using Thermo Scientific™ Dionex™ Viper™ Fingertight Fitting system * Get precise and accurate flow rates for stable retention times, robust peak integration, and confident identification from the Thermo Scientific™ Dionex™ SmartFlow™ technology
Chromeleon CDS Software Makes Operation Easy * Integrated service and validation monitoring with Thermo Scientific Dionex Chromeleon™ CDS, versions 6.x and 7.x * Automated system startup, standby and shutdown with Chromeleon CDS, versions 6.x and 7.x * Compatible with Chromeleon 7.x eWorkflow walk-up solution and advanced reporting features
Choose the Right Configuration for You * Four-solvent channel pump for versatile gradient methods, with integrated 4-channel degasser * (LPG-3400SD) or a single solvent channel pump for isocratic applications without degassing (ISO-3100SD) * Spacious solvent rack for six 1-liter solvent bottles or one large mobile phase reservoirs (SR-3000) * Thermostatted or nonthermostatted autosamplers/column compartments (ACC-3000T/ACC-3000) * Diode array, multiple wavelength, variable wavelength, fluorescence, charged aerosol, refractive index or mass spectrometry detectors * Fraction collector (AFC-3000) * Chromeleon 7.2 software package (standalone or networked, may also control your other chromatography equipment)
General Specifications
Detector | Variable Wavelength, UV |
Maximum Pressure | 9000 psi |
Standard Flow Rate | 0.2 to 10 ml/min |
Additional Specifications
PUMPS
Recommended Pump: LPG-3400SD
Flow Range: (Settable Range) 0.2–10 mL/min (0.001–10 mL/min)
Flow Accuracy: ±0.1%
Flow Precision: 0.9995, RSD <1% partial loop for 1–10 µL (20 µL loop)
Carryover: <0.02% for caffeine with external wash
Sample thermostating; With ACC-3000T: 8–40 °C, min. 15 °C below ambient
COLUMN COMPARTMENT
Recommended Column Compartment/Flow Manager: Integrated with ACC-3000
Pre-Heater Volume: Column inlet tubing
Temperature Range: 5 °C above ambient to 50 °C
DETECTOR
Recommended Detectors: VWD-3100
Detector Type: Variable Wavelength UV
Number of Channels: 1
Wavelength Range: 190–900 nm
Max. Data Collection Rate: 100 Hz
Recommended Flow Cell: Analytical: 11 µL, 10 mm path
Noise: <±3.5 µAU at 254 nm
Drift: <0.1 µAU/h