Thermo Scientific - UltiMate™ 3000 Basic Automated System

Manufactured by  Thermo Scientific
Follow this Equipment

Enjoy cost effective access to automated UHPLC separations in a simple, expandable system. The Thermo Scientific™...

Enjoy cost effective access to automated UHPLC separations in a simple, expandable system. The Thermo Scientific™ UltiMate™ 3000 Basic Automated system offers unique LC and entry level UHPLC capabilities without compromise. The integrated autosampler and thermostatted column compartment deliver retention time precision and the ability to fine tune backpressure, resolution and speed. With adjustable gradient delay volume, quaternary or isocratic pump, and excellent flow stability, the Automated Basic system is perfect for teaching or moderate sample throughput, routine applications.
Active Questions & AnswersAsk a Question

There are no current Discussions

UHPLC Systems Service ProvidersView All (7)

Request Support
Features of UltiMate™ 3000 Basic Automated System

Recommended Use  For moderate throughput laboratories that need LC and entry level UHPLC capabilities and are investing in their first systems. As an LC/UHPLC starter kit this automated modular system can be extended with additional detectors, a more spacious column compartment, or even access to column switching techniques. 

Typical application areas are teaching laboratories or quality control laboratories with moderate sample throughput requirements. The quaternary pump is recommended for methods deploying either conventional 3-5µm, or UHPLC sub-2µm particle columns where a moderately fast gradient response is acceptable for the benefit of increased mobile phase flexibility (i.e., multiple solvents, extended column wash, ternary or quaternary gradients). A four-channel solvent degasser is integrated in the pump module to simplify operation. 

Important Capabilities/Features: 

  • Up to 10 mL/min flow rate supporting up to 62 MPa (9000 psi/620 bar) backpressure for column format and column particle size flexibility * Fast and stable column thermostatting between 5°C above ambient and 50°C improves retention time precision. Compact and integrated autosampler column compartment to automate injections of multiple samples and calibration standards * Obtain near-zero dead volume for optimal peak shapes even over extended periods using Thermo Scientific™ Dionex™ Viper™ Fingertight Fitting system * Get precise and accurate flow rates for stable retention times, robust peak integration, and confident identification from the Thermo Scientific™ Dionex™ SmartFlow™ technology 

    Chromeleon CDS Software Makes Operation Easy * Integrated service and validation monitoring with Thermo Scientific Dionex Chromeleon™ CDS, versions 6.x and 7.x * Automated system startup, standby and shutdown with Chromeleon CDS, versions 6.x and 7.x * Compatible with Chromeleon 7.x eWorkflow walk-up solution and advanced reporting features 

    Choose the Right Configuration for You * Four-solvent channel pump for versatile gradient methods, with integrated 4-channel degasser * (LPG-3400SD) or a single solvent channel pump for isocratic applications without degassing (ISO-3100SD) * Spacious solvent rack for six 1-liter solvent bottles or one large mobile phase reservoirs (SR-3000) * Thermostatted or nonthermostatted autosamplers/column compartments (ACC-3000T/ACC-3000) * Diode array, multiple wavelength, variable wavelength, fluorescence, charged aerosol, refractive index or mass spectrometry detectors * Fraction collector (AFC-3000) * Chromeleon 7.2 software package (standalone or networked, may also control your other chromatography equipment)

General Specifications
DetectorVariable Wavelength, UV
Maximum Pressure9000 psi
Standard Flow Rate 0.2 to 10 ml/min
Additional Specifications

PUMPS    
    
Recommended Pump: LPG-3400SD    
Flow Range: (Settable Range) 0.2–10 mL/min (0.001–10 mL/min)    
Flow Accuracy: ±0.1%    
Flow Precision: 0.9995, RSD <1% partial loop for 1–10 µL (20 µL loop)    
Carryover: <0.02% for caffeine with external wash    
Sample thermostating; With ACC-3000T: 8–40 °C, min. 15 °C below ambient    
    
COLUMN COMPARTMENT    
    
Recommended Column Compartment/Flow Manager: Integrated with ACC-3000    
Pre-Heater Volume: Column inlet tubing    
Temperature Range: 5 °C above ambient to 50 °C    
    
DETECTOR    
    
Recommended Detectors: VWD-3100    
Detector Type: Variable Wavelength UV    
Number of Channels: 1    
Wavelength Range: 190–900 nm    
Max. Data Collection Rate: 100 Hz    
Recommended Flow Cell: Analytical: 11 µL, 10 mm path    
Noise: <±3.5 µAU at 254 nm    
Drift: <0.1 µAU/h

Looking for New or Used Equipment?

Shop on LabX

Shop on Labx.com