Amerex Instruments - HA-300MD
Self-contained, portable top-loading autoclave with heating drying cycle
- Top loading design is ideal for the sterilization of media and other liquids in tall flasks and bottles.
- Deep chamber for accommodating fermenters and bioreactors.
- Space-saving, stand-alone unit occupies minimal floor space and can fit conveniently into many locations in the laboratory.
- Full-count sterilization guarantees sterilization at the preset temperature and time. Timer pauses if chamber temperature falls 1ºC below the preset temperature
- Liquid cycles with slow exhaust or natural cooling for liquids and solids without drying.
- Drying cycle for glassware and other solid loads.
- Additional drying cycle without re-sterilization.
- Micro-processor controlled for precise sterilization.
- Internal steam generator uses purified water to generate pure steam.
- Audio/visual alarms and on-board diagnostics for rapid repair.
Active Questions & AnswersAsk a Question
There are no current Discussions
Autoclaves Service ProvidersView All (25)
Documents & Manuals
There are no Documents or Manuals available.
Features
There are no Features available.
General Specifications
Depth | 20.5 in |
Height | 42.1 in |
Width | 17.3 in |
Power Requirements | 120V, 50/60 Hz, single phase |
Orientation | Floor Standing |
Peak Power | 2000 W |
Chamber Volume | 50 L |
Weight | 180 lbs (shipping) |
Additional Specifications
TEMPERATURE
Display Range: 5ºC to 129ºC
GAUGE PRESSURE
Operating Range (gauge): 3 to 30 psi (gauge), 0.021 to 0.21MPa
TIMER
Sterilization: 1 to 999 minutes, or continuous
Drying: 1 to 120 minutes
SAFETY DEVICES Over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, over-current detector and circuit breaker
PROGRAMS
1. For solids with drying: Sterilization -> Manual Steam/H2O Exhaust -> Drying -> Completion
2. For solids w/o drying:Sterilization -> Auto Steam Exhaust -> Completion
3. For liquids: Sterilization -> Natural Cooling -> Completion
4. For additional drying: Drying -> Completion
ACCESSORIES
Stainless steel: Mesh wire baskets, buckets (pails), and test tube rack