Features of UltiMate 3000 RSLCnano System
Recommended Use:
For all proteomics labs looking for the highest performance and usability in nano LC/MS. * It is easily configured as routine nano LC instrument for LC/MS protein identification. * The versatility and excellent performance allow users to optimize sample throughput and protein identification. * Routine, advanced and high throughput proteomics applications, like direct-injection, pre-concentration, 2D-LC techniques or tandem operation, are all supported by standardized workflows. * Customized application setups are easily established for proteomics, metabolomics, intact protein analysis, biopharma and beyond. * The system’s dual pumps cover a very wide flow range starting at the very low flow rates from 20nL/min. less than 50µm i.d. to 1mm column i.d. flows up to 50µL/min. with the nano pump. * The loading pump can deliver accurate gradient from 50 to 2500µL/min. for 2D separations or an occasional analytical flow experiment.
Exceptional Chromatography Performance: The UltiMate 3000 RSCLnano system delivers the highest level of performance for your low flow applications: * Up to three times faster than conventional nano LC systems * Up to twice the resolution of traditional nano LC systems * Extensive flow-pressure range for maximum application flexibility * Continuous direct flow delivery enables uninterrupted, very long ultra-high resolution runs * Optional stainless-steel-free column flow path protects your analytes from modification and degradation
Extended Solution to Match Your Workflows: * Extensive selection of column formats, traps and chemistries, e.g., 25cm and 50cm long 75µm I.D. Thermo Scientific™ Acclaim™ PepMap™ columns * Complete MS integration with Thermo Scientific Dionex™ DCMSLink™ software Integrates seamlessly with Thermo Scientific™ EASY-Spray™ * Full range of consumables for all application areas, including nanoViper-based application kits
Choose the Right Configuration for You: * Nano pump, loading pump and heated column compartment integrated in one setup * Compact separate and heated column compartment featuring a direct nano/cap/micro flow, binary pump and an exceptionally capable loading pump (NCS-3500) * Optional two-position, six-port or two-position, ten-port Snap-In nano valves (also PEAK versions) * Four-channel degasser built into the solvent rack (SRD-3400) * Thermostatted pull-loop well plate autosampler (WPS-3000TPLRS) or thermostatted fractionation sampler for advanced 2D-LC techniques (WPS-3000TFC) * Variable wavelength UV detector with large choice of low volume U-Z view flow cells (e.g., 3nL) * nanoViper MS connection kits for Thermo Scientific mass spectrometers * Optional EASY-Spray source interfaces * Optional additional nano/cap/micro flow pump module (NCP-3200) * DCMSLink system control integration software
General Specifications
Injection Volume | 0.01 to 125 μL |
Pump Flow Rate | 0.02 to 50 µL/min |
UV/VIS Noise | 0.00005 AU at 254 nm |
UV/VIS Wavelength | 190 to 900 nm |
Additional Specifications
Pump
Flow Rate Ranges: HPG Nano Pump: 20 nL/min–50 ΜL/min (using dedicated flow selectors)
Flow Selectors (recommended [settable]):
Nano: 50 nL/min–1000 nL/min
[1 nL/min–1500 nL/min]
Capillary: 500 nL/min–10 ΜL/min
[1 nL/min–15 ΜL/min]
Micro: 2.5 ΜL/min–50 ΜL/min
[1 nL/min–50 ΜL/min]
Custom flow selectors are available on request.
LPG Micro Pump: 10–2500 ΜL/min (NCS-3500RS only)
Pressure Range:
20–800 bar (300–11,600 psi)
20–500 bar (300–7250 psi), loading pump
Number of Solvent Channels:
HPG Nano Pump: 2
LPG Micro Pump: 3
Flow Calibration: Semi-automated
Proportioning Accuracy: 0.9995 from 100–500 nL injections
Carryover: <0.02% with needle wash (caffeine)
Optional Sample Cooling: 4 °C–45 °C, at least 22 °C below ambient
Fraction Collection: Micro fraction collection option [up to 345 bar (5000 psi)]
Wetted Parts: PEEK, Stainless steel, PCTFE, fused silica
Dimensions (h × w × d): 16 × 16.5 × 20 in.
Weight: 22.7 kg (50 lb) including cooling
Power Requirements: 85–260 V, 50/60 Hz, max. 320 W
PC Connection: USB; USB hub with three integrated sockets
I/O Interfaces: Four digital inputs and four programmable outputs
Detector
Data Collection Rate: Up to 200 Hz (in single wavelength mode)
Maximum Number of Channels: Four
Drift: 4.0 mAU/h
Wavelength Range: 190–900 nm
Noise: Typically <0.05 mAU at 254 nm
Lamp: Deuterium lamp, Tungsten lamp
Flow Cell Volume: 3 nL for nano LC; 45 nL for capillary LC; 180 nL for micro LC
Dimensions (h × w × d): 16 × 42 × 51 cm (6.3 × 16.5 × 20 in.)
Weight: 15 kg (33 lb)
Power Requirements: 85–260 V, 50/60 Hz, max. 150 W
PC Connection: USB
I/O Interfaces:
Four digital inputs and four digital outputs
Two analog inputs (optional DAC module)