PerkinElmer - DSC 6000

Manufactured by  PerkinElmer
Follow this Equipment

See what enhanced performance can do for you. DSC 6000 – Advanced single furnace design with MT-DSC

See what enhanced performance can do for you. DSC 6000 – Advanced single furnace design with MT-DSC   
  
Our DSC 6000 gives you all the advantages of the DSC 4000 and more. Designed for research application, the DSC 6000 comes with Modulated Temperature DSC (MT-DSC) technology for easier data interpretation, and new capabilities for product development and troubleshooting. The DSC can also be fitted with an optional autosampler and photocalorimetry accessory.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Updated byOkan

Need Equipment Support?

Request Support
Features of DSC 6000
  • MT-DSC, enabling you to separate kinetic events and interpret data with greater ease   
  • Optional photocalorimetry accessory for more advanced analysis of photo-initiated reactions  
  • Scanning rates up to 0.1 ºC to 100 ºC for broad applications flexibility  

Typical applications for the DSC 6000:   

  • Characterizing materials during product and process development  
  • Trouble-shooting product quality issues   
  • Academic research and post-graduate studies
General Specifications
Cooling Rate0.1 to 100 °C/min
Depth 16.5 in
DSC TypeHeat Flux DSC
Height6.7 in
Temperature Range-180 to 450 °C
Width 15 in
Heating Rate0.1 to 100 °C/min
Power Requirements 100-240 Volt 50/60 Hz
Weight26 lbs to 36lbs
Additional Specifications

Calorimetric Performance 
Dynamic range: ±175 mW  
Accuracy: ±2% 
Precision: ±0.1% 
Indium height/width (mW/°C):
Indium melting time (sec): 3.3 
Digital resolution: 0.02 ΜW 
 
Temperature Performance 
Accuracy: ±0.1 °C 
Precision: ±0.02 °C 
Data points/sec: 10 
Cooling times – intracooler (100 °C to 0 °C): Under 4 minutes

Looking for

PerkinElmer - DSC 6000?

Shop on LabX.com