Netzsch - DSC 204 F1 Phoenix®
Manufactured by Netzsch
The premium DSC for all requirements
The versatile Premium Differential Scanning Calorimeter, DSC 204 F1 Phoenix®, fascinates with its unique concept: all essential operational components are integrated into the instrument and optional add-ons can be installed anytime on site.
The DSC measuring cell consists of a cylindrical high-conductivity silver block with an embedded heating coil for broad thermal symmetry (3D symmetry) in the sample chamber, the cooling ports for liquid nitrogen or compressed air cooling and a cooling ring for connection of the intracooler (also with simultaneous liquid nitrogen cooling). The gas-tight construction and integrated mass flow controllers for the purge and protective gases allow coupling to an FTIR or MS for gas analysis.
The DSC measuring cell consists of a cylindrical high-conductivity silver block with an embedded heating coil for broad thermal symmetry (3D symmetry) in the sample chamber, the cooling ports for liquid nitrogen or compressed air cooling and a cooling ring for connection of the intracooler (also with simultaneous liquid nitrogen cooling). The gas-tight construction and integrated mass flow controllers for the purge and protective gases allow coupling to an FTIR or MS for gas analysis.
Active Questions & AnswersAsk a Question
There are no current Discussions
Need Equipment Support?
Documents & ManualsView All Documents
Features of DSC 204 F1 Phoenix®
- Very short time constant with T-Sensor
- Extremely high sensitivity with µ-Sensor, application example
- Baseline optimized by BeFlat®
- Gas-tight design
- Gas control for 2 sample gases and 1 protective gas by integrated mass flow control system and software
- Compressed air cooling 25 to 700 °C
- Improved intra cooler -85 to 600 °C
- Fast controlled liquid nitrogen cooler -180 to 700 °C
- Automatic sample changer: ASC for 64 sample and reference crucibles (Option)
- Macro Recorder for routine evaluations (Option)
General Specifications
| Cooling Rate | 0.001 to 200 K/min |
| Temperature Range | 25 to 700 °C |
| Heating Rate | 0.001 to 200 K/min |
Additional Specifications
Temperature range
with air cooling:
-85°C to 600°C
with Intracooler:
-180°C to 700°C
Heating rates: 0.001 K/min to 200 K/min
Cooling rates: 0.001 K/min to 200 K/min (depending upon the cooling system and the temperature range)
DSC limit of detection: < 0.1 ΜW (depending upon the sensor)
System time constant: 0.6 s to 3 s (depending upon the sensor)