The push-plug optics of the D8 ADVANCE enables a quick switch between the para-focusing Bragg-Brentano geometry to parallel beam geometry without realignment of the system. For the “cut above” standard powder diffraction, the D8 ADVANCE with Vario1 Johansson focusing monochromator features high-flux K-alpha-1 radiation for reflection as well as for foil or capillary transmission experiments.
Active Questions & AnswersAsk a Question
connection problem
Recent Questions & Answers
Asked byAyoub GROULI
X-ray Diffraction Service ProvidersView All (9)
Documents & ManualsView All Documents
Features of D8 ADVANCE
D8 ADVANCE - Sample stage concept
The sample stage concept of the D8 ADVANCE allows for many different sample environments. These range from 10 K Helium closed-cycle cryostats up to non-ambient sample conditions with high temperature and humidity. For ambient conditions, the flip-stick sample stage and the auto changer provide the capability of measuring a batch of up to 90 samples in transmission as well as in reflection geometry.
Detector technology is an important core competence of Bruker AXS. This means that the D8 ADVANCE can be equipped with the full range of point and linear detectors. According to the analytical requirements, the user can choose between different types of dynamic scintillation counter, energy dispersive SOL-XE, or the Super Speed VÅNTEC-1.
No matter how the individual diffraction solution is configured, any D8 ADVANCE has the ability to perform a full range of applications, from qualitative and quantitative phase identification, under ambient or non ambient conditions, to crystal structure solution from powder samples, crystallite size determination, micro strain analysis, residual stress analysis, and preferred orientation.
General Specifications
Depth | 1135 mm |
Height | 1868 mm |
Width | 1300 mm |
Weight | 770 kg |
Additional Specifications
Configurations:
Vertical goniometer, Theta/2Theta or Theta/Theta geometry
Measuring circle diameter:
Predefined at 500 mm, 560 mm and 600 mm or any intermediate setting
Angular range:
360°
Max. usable angular range:
–110° < 2Theta 168°
Angle positioning:
Stepper motors with optical encoders
Smallest addressable increment:
0.0001°
Instrument alignment:
Equal or better than ± 0.01° 2Theta; NIST SRM 1976a always included
Maximum angular speed:
20°/s
Detectors:
Point detectors:
Scintillation counter
SOL-XE energy dispersive
1-D detectors:
LYNXEYE
VÅNTEC-1
All detectors guaranteed without defective/dead strips or areas
Cooling water supply:
Min. 4 l/min, pressure 4 bar to 7.5 bar, no pressure on outlet side,
temperature: 10 °C to 20 °C
Power supply:
Single phase: 208 to 240 V
Three phases: 120 V, 230 V, 240 V;
47 to 63 Hz