Shimadzu - SIL-HTa/SIL-HTc

Manufactured by  Shimadzu
Follow this Equipment

SIL-HTa/SIL-HTc

High throughput is the key word for modern HPLC analysis. The SIL-HT is the best performing autosampler in terms of sample injection speed, sample capacity, and minimizing sample carryover of highly adsorbent sample components. These features are in increasingly high demand and essential for high throughput analysis. The SIL-HT has been developed to meet such users' requirements and will also provide optimum results in your lab. The SIL-HT is available in two versions: the SIL-HTA, ambient model; and the SIL-HTC with sample temperature control from 4 to 40C.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Asked byDennis Yasuda

Autosamplers Service ProvidersView All (1)

Request Support
Features of SIL-HTa/SIL-HTc

High throughput     The SIL-HT maximizes analysis throughput with fast sample injection and high sample capacity. Injection speed is 15 seconds for injecting samples of 10uL .    

Near-zero sample carryover     Recently, an increasing number of extremely basic and hydrophobic components are being used in new drug formulations. Since these chemical compounds strongly adsorb to the sampling needles or rotor seals (mainly Vespel) of autosamplers, cross contamination may not be reduced by simple rinsing. To minimize adsorption of such compounds, the outer surface of the sampling needle of the SIL-HT's autosampler is coated using a special, innovative surface processing technology (patent pending). Furthermore, by employing newly developed rotor seals and needle seals made of PEEK, the SIL-HT dramatically reduces contamination, even with the most highly adsorbent sample compounds. As a result, the SIL-HT demonstrates nearly zero sample carryover.          Excellent repeatability     The SIL-HT uses a needle-in-the-flow-path injection design (also called direct injection method or total volume injection method) and a high-performance measuring pump (6nL step resolution) for enhanced accuracy. This results in superior repeatability over conventional autosamplers. In addition, precious samples are not wasted since the total volume aspirated is injected.

General Specifications
Cross Contamination0.01 % (for analysis of naphthalene)
Depth 500 mm
Height415 mm
Maximum Pressure20 MPa
Width 540 mm
Power Requirements AC100V, 150VA (SIL-HTa), 300VA (SIL-HTc), 50/60Hz
Optional Injection Range1 to 2000 μL
Standard Injection Range0.1 to 100 μL
Weight40 kg (SIL-HTa), 45kg (SIL-HTc)
Additional Specifications

Sample Injection Unit    
Sample Injection System: Total sample injection, variable sample injection system (zero sample loss during injection)    
Sample Processing Capability:     
With 1-mL vials: 350    
With 1.50mL vials: 210(SIL-HTa), 140(SIL-HTc)    
With 4-mL vials:100    
With micro plates:4 (96 wells: 384, 384 wells: 1536)    
With deep-well plates: 4 (96 wells: 384, 384 wells: 1536)    
Injection Volume Repeatability: RSD: 0.3% max. (at 10µL injection)    
Injection Volume Accuracy:     
Less than +/-1% (50µL)    
Injection Volume Accuracy Correction:     
Possible    
Number Of Injections:     
1-99 times per sample    
Settting Of Analysis time:     
0.01-9999.9    
(0.01 minute increments to 999.099min.,0.1 minute increments over 1000min.)    
Number Of Steps For Injection Conditions:     
Max 102 steps    
Sample Suction Speed:     
0.1 - 15µL/sec (0.1µL/s step for 0.1-0.9µL/sec., 1µL/s step for over 1µL/sec.)    
Rinsing Liquid Suction Speed:     
1 - 35µL/sec (1µL/s step)    
    
Operating pH Range:     
pH: 1 to 14 standard    
Wetted Surface Material:     
SUS316, PEEK, Ceramics    
Operating Temperature Range:     
4 to 35°C    
Operating Humidity Range:     
up to 70%