BINDER - KB series
Manufactured by BINDER
Refrigerated incubators with forced convection
The most versatile incubator for microorganisms: The BINDER KB series controls a temperature range of -5 °C to 100 °C. With its comprehensive program functions, it offers a wide range of capabilities and delivers reproducible test results.
Active Questions & AnswersAsk a Question
There are no current Discussions
Refrigerated Incubator Service ProvidersView All (28)
Documents & ManualsView All Documents
Features of KB series
- Electronically controlled APT.line™ preheating chamber with DCT™ cooling system assuring temperature accuracy and reproducible results
- Temperature range 0 °C to 100 °C
- MP controller with 2 programs with 10 sections each, alternatively switchable to 1 program with 20 sections
- Integrated weekly program timer with real-time function -Adjustable ramp function via program editor -Digital temperature setting with an accuracy of a tenth of a degree -Adjustable fan speed -Elapsed time indicator
- Independent temperature safety device class 3.1 (DIN 12880) with optical and audible temperature alarm
- Inner glass door
- RS 422 interface for use with APT-COM™ DataControlSystem communication software or switch over to printer output with RS 232 / RS 422 interface converter
- Adjustable intervals for printer
- Units up to 115 liters are stackable
- 2 shelves, stainless steel
- BINDER test certificate
General Specifications
There are no General Specifications available.
Series Modals
BINDER - KB 115
[+] View Model Specifications| Depth | 646 mm (plus door handle, I-panel and connection 64 mm) |
| Height | 1022 mm |
| Width | 834 mm |
| Weight | 105 kg |
| Additional Model Specifications | [u]Interior dimensions[/u]
Width (mm): 600 Height (mm): 480 Depth (mm): 400 Interior volume (l): 115 Racks (number standard/max.): 2 / 5 Load per rack (kg): 20 Permitted total load (kg): 50 [u]Temperature data[/u] Temperature range (°C) 1): -5 - 100 Temperature variation: at 4 °C (± K): 0,4 at 25 °C (± K): 0,1 at 37 °C (± K): 0,2 Temperature fluctuation max. (± K): 0,1 Recovery time after door was opened for 30 sec 2): at 4 °C (min.): 5 at 37 °C (min.): 1 [u]Electrical data[/u] IP protection class acc. to EN 60529: IP 20 Voltage (± 10%) 50 / 60 Hz (V): 230 1N~ Nominal power (kW): 0,46 Energy consumption 3): at 37 °C (W): 115 Noise level (dB (A):) 47 |
BINDER - KB 23
[+] View Model Specifications| Depth | 516 mm (plus door handle, I-panel and connection 64 mm) |
| Height | 618 mm |
| Width | 433 mm |
| Weight | 44 kg |
| Additional Model Specifications | [u]Interior dimensions[/u]
Width (mm): 222 Height (mm): 330 Depth (mm): 277 Interior volume (l): 20 Racks (number standard/max.): 2 / 3 Load per rack (kg): 10 Permitted total load (kg): 25 [u]Temperature data[/u] Temperature range (°C) 1): 0 - 100 Temperature variation: at 4 °C (± K): 0,7 at 25 °C (± K): 0,3 at 37 °C (± K): 0,3 Temperature fluctuation max. (± K): 0,2 Recovery time after door was opened for 30 sec 2): at 4 °C (min.): 5 at 37 °C (min.): 2 [u]Electrical data[/u] IP protection class acc. to EN 60529: IP 20 Voltage (± 10%) 50 / 60 Hz (V): 230 1N~ Nominal power (kW): 0,34 Energy consumption 3): at 37 °C (W): 60 Noise level (dB (A):) 47 |
BINDER - KB 240
[+] View Model Specifications| Depth | 800 mm |
| Height | 1460 mm |
| Width | 925 mm |
| Weight | 170 kg |
| Additional Model Specifications | [u]Interior dimensions[/u]
Width (mm): 650 Height (mm): 785 Depth (mm): 485 Interior volume (l): 247 Racks (number standard/max.): 2 / 9 Load per rack (kg): 30 Permitted total load (kg): 100 [u]Temperature data[/u] Temperature range (°C) 1): -5 - 100 Temperature variation: at 4 °C (± K): 0,5 at 25 °C (± K): 0,2 at 37 °C (± K): 0,2 Temperature fluctuation max. (± K): 0,1 Recovery time after door was opened for 30 sec 2): at 4 °C (min.): 16 at 37 °C (min.): 1 [u]Electrical data[/u] IP protection class acc. to EN 60529: IP 20 Voltage (± 10%) 50 / 60 Hz (V): 200-240 1N~ Nominal power (kW): 1,2 Energy consumption 3): at 37 °C (W): 260 Noise level (dB (A):) 52 |
BINDER - KB 400
[+] View Model Specifications| Depth | 800 mm (plus door handle, I-panel and connection 50 mm) |
| Height | 1945 mm |
| Width | 925 mm |
| Weight | 220 kg |
| Additional Model Specifications | [u]Interior dimensions[/u]
Width (mm): 650 Height (mm): 1270 Depth (mm): 485 Interior volume (l): 400 Racks (number standard/max.): 2 / 15 Load per rack (kg): 30 Permitted total load (kg): 100 [u]Temperature data[/u] Temperature range (°C) 1): -5 - 100 Temperature variation max.(± K): 0,6 at 4 °C (± K): 0,4 at 25 °C (± K): 0,2 at 37 °C (± K): 0,3 Temperature fluctuation max. (± K): 0,1 Recovery time after door was opened for 30 sec 2): at 4 °C (min.): 14 at 37 °C (min.): 3 [u]Electrical data[/u] IP protection class acc. to EN 60529: IP 20 Voltage (± 10%) 50 / 60 Hz (V): 200-240 1N~ Nominal power (kW): 1,4 Energy consumption 3): at 37 °C (W): 420 Noise level (dB (A):) 53 |
BINDER - KB 53
[+] View Model Specifications| Depth | 576 mm (plus door handle, I-panel and connection 64 mm) |
| Height | 837 mm |
| Width | 634 mm |
| Weight | 72 kg |
| Additional Model Specifications | [u]Interior dimensions[/u]
Width (mm): 400 Height (mm): 400 Depth (mm): 330 Interior volume (l): 53 Racks (number standard/max.): 2 / 4 Load per rack (kg): 15 Permitted total load (kg): 40 [u]Temperature data[/u] Temperature range (°C) 1): -5 - 100 Temperature variation: at 4 °C (± K): 0,6 at 25 °C (± K): 0,2 at 37 °C (± K): 0,3 Temperature fluctuation max. (± K): 0,1 Recovery time after door was opened for 30 sec 2): at 4 °C (min.): 4 at 37 °C (min.): 1 [u]Electrical data[/u] IP protection class acc. to EN 60529: IP 20 Voltage (± 10%) 50 / 60 Hz (V): 230 1N~ Nominal power (kW): 0,46 Energy consumption 3): at 37 °C (W): 215 Noise level (dB (A):) 47 |
BINDER - KB 720
[+] View Model Specifications| Depth | 887 mm (plus door handle, I-panel and connection 52 mm) |
| Height | 1925 mm |
| Width | 1250 mm |
| Weight | 309 kg |
| Additional Model Specifications | [u]Interior dimensions[/u]
Width (mm): 970 Height (mm): 1250 Depth (mm): 576 Interior volume (l): 698 Racks (number standard/max.): 2 / 15 Load per rack (kg): 45 Permitted total load (kg): 150 [u]Temperature data[/u] Temperature range (°C) 1): -5 - 100 Temperature variation max.(± C): 0,5 at 4 °C (± K): 0,2 at 25 °C (± K): 0,2 at 37 °C (± K): 0,2 Temperature fluctuation max. (± K): 0,1 Recovery time after door was opened for 30 sec 2): at 4 °C (min.): 12 at 37 °C (min.): 2 [u]Electrical data[/u] IP protection class acc. to EN 60529: IP 20 Voltage (± 10%) 50 / 60 Hz (V): 200-240 1N~ Nominal power (kW): 2,3 Energy consumption 3): at 37 °C (W): 510 Noise level (dB (A):) 53 |