Denver Instrument - IR-60

Manufactured by  Denver Instrument
Follow this Equipment

Big features in a small footprint.

Rapid heating technology and easy setup are combined in a rugged, straightforward instrument. The perfect solution for routine analysis of up to 5 unique samples!

New heating technology.

For over a decade Denver Instrument has provided superior infrared moisture analysis with quartz heating elements. The IR-60 utilizes a new quartz design. The coil configuration allows rapid heating, similar to halogen, while assuring even heating of the sample.

Active Questions & AnswersAsk a Question

There are no current Discussions

Moisture Analyzers Service ProvidersView All (3)

Request Support
Features of IR-60

Flexible endpoint determination

For general applications, choose the factory-defined automatic endpoint or timer mode. Or tailor the endpoint specifically for your samples to get better repeatability or shorter analysis time.

Setup made simple

Store up to 5 programs for analysis of different sample types. Standby temperature keeps the instrument ready for the next test and reduces analysis time. Password protection blocks unauthorized changes.

Select a program and go

Place an empty pan on the instrument and press the Enter key. Add sample, close the hood and the testing will begin automatically. Display shows test parameters, current temperature, elapsed test time and current sample value.

  • Efficient use of space, only 12.6"L x 8.4"W x 7.1"H
  • Quartz coil infrared heating element
  • Single-block weighing technology with 0.001g resolution for precise results
  • Readability of 0.01%
  • Store up to 5 programs
  • Factory-defined or semi-adjustable endpoint criteria
  • RS232 serial port to send information to computer or optional printer
General Specifications
Depth 12.6 in
Height7.1 in
Width 8.4 in
Power Requirements 115 VAC , 50/60 Hz
Weight12.1 lbs
Additional Specifications

Working Range: 0.1 - 99.9%
Measurement Method: Infrared radiation, loss on drying
Heat Source: Quartz coil
Temperature Setting: 40° - 210°C in one degree increments
Temperature Control: Temperature sensor, +/- 5°C
Temperature Steps: One
End of Analysis Mode: Time-out, automatic slope or semi-adjustable slope
Readability of Results: 0.01%
Balance Capability: 60 grams
Balance Readability: 1mg (0.001g)
Program Storage: 5 programs
Data Storage: Current result

Looking for New or Used Equipment?

Shop on LabX