Labconco - XPert Weigh Boxes
Manufactured by Labconco
Has inlet and outlet HEPA filters and provide a leak-tight physical barrier to protect the operator from hazardous airborne particulates and powders.
XPert Weigh Boxes have inlet and outlet HEPA filters and provide a leak-tight physical barrier to protect the operator from hazardous airborne particulates and powders. These boxes include an anti-static ionizer fan to reduce static charge and a balance vibration isolator, both useful during weighing operations.
Active Questions & AnswersAsk a Question
There are no current Discussions
Glove Boxes Service ProvidersView All (2)
Documents & Manuals
There are no Documents or Manuals available.
Features of XPert Weigh Boxes
- One-piece molded, medium-density polyethylene shell. Labconco recommends this website for a comprehensive listing of chemical resistance data for polyethylene.
- Dry powder epoxy-coated steel superstructure
- 1/4" thick laminated safety glass viewing window, 26.4" wide x 12.5" high, removable for loading equipment, with powder-coated steel frame and neoprene gasket
- Space-saving inner and outer transfer chamber doors that pivot upward, are counterbalanced and equipped with quick-latches
- Black, chemical-resistant work surface sealed to the glove box floor
- Anti-static ionizer fan to circulate ionized air inside the box to reduce static charge during weighing
- Interior ceiling-mounted electrical duplex receptacle with 5 amp maximum circuit load
- Neoprene gloves, size 9 3/4", are included.
- Built-in blower with speed control capable of airflow from 13 CFM to 100 CFM.
- Fluorescent lamp, 22 watts, with easy replacement from the top
- Switches for light, blower and innermost interior electrical receptacle
- Inlet and outlet HEPA filters, 99.99% efficient on 0.3 micron particulates
- Test port for challenging HEPA filter integrity
- Main chamber static pressure gauge for monitoring HEPA filter loading.
- Approximately 13 cubic feet of interior space in main chamber
- 11.0" nominal ID x 12.0" long transfer chamber
- Factory leak tested with a mass spectrometer while glove box is pressurized with helium at 5 inches of water gauge
- No detectable leaks greater than 1 x 10(-3) ml/sec
- Meet Class 1 atmosphere containment conditions for oxygen leak rates per ISO 10648-2 test method
- Transfer chamber interior interlocking shelf of dry powder epoxy-coated steel
- 0.05 ppm average tracer gas levels or less detected during perimeter scan testing (surpasses ASHRAE 110 standards)
- Airborne particulate cleanliness (measured under positive and negative pressures per ISO 14644-1 test method) exceeds ISO Class 5 conditions (equivalent to Class 100) and achieved ISO Class 3 conditions (equivalent to Class 1) at both minimum and maximum operational airflow volumes
- Vibration less than 1.5 x 10(-5) inches displacement measured at the work surface at both minimum and maximum airflow volumes
- ETL listed (115 volt models)
- CE Conformity (230 volt models)
General Specifications
Depth | 31.6 in |
Height | 40 in |
Width | 52.7 in |
Internal Width | 33.5 in |
Power Requirements | 115 volts, 60 Hz, 8 amps |
Number of Glove Ports | 2 |
Glove Box Construction | Polyethylene |
Glove Box Type | Stand Alone |
Ducting | Ductless |
Additional Specifications
Filter Type: HEPA
Primary Application: Work with hazardous materials
Electrical Requirements: Domestic
Power Cord & Plug: 115 volts, 8 amps
Required Accessories: supporting base
Built-In Options: None
Compliance: ETL listing, UL 3101-1/61010-1 and CAN/CSA C22.2 No. 1010.1 conformity
Interior-mounted electrical outlets: 115 volt
Note: Also available for other electrical requirements