Environics - Series 4000

Manufactured by  Environics
Follow this Equipment

Multi-Component Gas Mixing System

The Environics® Series 4000 is a multi-component gas mixing system that automatically mixes up to three individual gases in a balance gas. The gas mixes can be used in generating precise gas calibration standards, creating gaseous atmospheres or producing gas mixes for analytical research or production purposes. The Series 4000 can produce gas concentrations from percent to ppb levels for single or multi-point calibration.      
      
The system consists of two components: the Series 4000 instrument and the user's personal computer. The user interface is a Microsoft® Windows application that communicates with the Environics system via an RS232 serial interface.      
      
The Series 4000 consists of a single chassis supporting up to four mass flow controllers. The mass flow controllers are factory calibrated using a primary flow standard traceable to the United States National Institute of Standards and Technology (NIST). Each flow controller utilizes an 11 point calibration table with linear interpolation, to increase accuracy and reduce flow controller nonlinearity.
Active Questions & AnswersAsk a Question

There are no current Discussions

Gas and Air Analyzers Service ProvidersView All (10)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of Series 4000
  • Simultaneously mixes up to three gases in a balance gas at user set concentrations.      
  • Allows multi-point calibration of analyzers.      
  • Automatic calculation of dilution and span gas flows.      
  • Internally-stored Mass Flow Controller calibration tables .      
  • Modular design allows for additional gas circuits to be added.      
  • User definable cylinder library allows for easy selection of frequently used gas cylinders. Each cylinder may contain an unlimited number of component gases, with automatic K-factor calculation.      

INSTRUMENT OPERATION   

The Series 4000 has three basic modes of operation.  

  • Concentration Mode   

Allows user to create a blend by entering target gas concentrations for each source gas cylinder, and the desired total output flow for the mix.  

  • Flow Mode   

Allows user to specify the flow rate of each gas cylinder.  

  • Program/Schedule Modes   

Provides the ability to program the instrument for unattended operation. Programs can be recalled and run in any sequence, at various times/dates.

General Specifications
Depth 15 in
Height7 in
Width 17 in
Power Requirements 110 to 240 VAC, 50/60 Hz
Additional Specifications

Performance      
Accuracy      
Concentration: ± 1.0% setpoint      
Flow: ± 1.0% setpoint      
Repeatability: 0.05%      
Warm up time: 30 minutes       
      
Mechanical      
Gas connections: 1/4" Swagelok (or compatible fitting)      
Operating Pressures      
Minimum: 10 psig (0.67 Bar)      
Recommended: 25 psig (1.68 Bar)      
Maximum: 75 psig (5.04 Bar)      
      
Wetted Surfaces      
Tubing: Electropolished 316 Stainless Steel      
MFCs: Stainless Steel      
Seals: Viton®      
(Optional: Kalrez®, Buna-N, Neoprene, Metal)      
Operating temperatures|:      
32° - 104° F (0° - 40° C)