Environics - Series 2014
Computerized VOC Gas Dilution System
EPA methods for VOC's specify a dynamic dilution to produce the initial multi-point calibration as well as the required daily and pre-sample calibration at the nominal 10 ppb level.
The Series 2014 employs mass flow controllers calibrated to a NIST (National Institute of Standards and Technology) traceable primary standard. All gas wetted surfaces are electropolished stainless steel to minimize absorption effects at ppb levels. Seals are gas compatible elastomers. A heated flow path is provided.
Commands are entered from the front panel and displayed on a backlit 25 line by 80 character liquid crystal display. Calibration sequences are stored in the internal microprocessor for recall by the keypad, optional RS-232 communications or optional status interface.
Humidification and pressurization options are available.
A purge circuit allows for nitrogen purge of the gas wetted surfaces to minimize contamination.
The Series 2014 is available in either a standard 19" rack mount or bench top configuration
Active Questions & AnswersAsk a Question
There are no current Discussions
Gas and Air Analyzers Service ProvidersView All (10)
Documents & Manuals
There are no Documents or Manuals available.
Features of Series 2014
- Dynamic mixing with automatic computation of component and balance gas flow rates allows the user to formulate ppb standards and vary the concentration.
- Minimizes inventory of expensive, long lead time, VOC gas cylinders. Save on gas costs, cylinder rental and handling labor.
- Computerized K-factor correction insures accurate use of all gases based on thermal density.
- Internally-stored mass flow controller calibration data improves accuracy.
- High capacity memory permits storage and recall of up to 200 multi-component "recipes", saving time and reducing errors.
- 25 line by 80 character display permits viewing of data in worksheet form.
- Modular design allows user to add additional gas circuits later, reducing initial investment and protecting against obsolescence.
- Optional RS-232 Serial Data Interface permits remote operation and complete integration with a data station.
- Optional Status board allows user to remotely activate different modes of the system and also activate external devices.
SOFTWARE
The Series 2014 has six primary routines accessible through on screen "soft" keys.
- Concentration Mode: User enters target output gas concentration for the span gas. The actual concentration is displayed during mixing.
- Flow Mode: User enters target flow rate for each component gas. Actual flow rates are displayed after mixing is initiated.
- Maintain Ports: User enters the name of the component gas in the source cylinder, its concentration and the port to which it is connected.
- Divider Mode: Allows the user to operate the instrument as a computerized ten step gas divider.
- Automatic Sequencer: Permits unattended automatic operation of the instrument on a programmable seven-day schedule.
- Purge Mode: Purge component gas circuits and mixing zone.
General Specifications
Depth | 23.5 in |
Height | 7 in |
Width | 17 in |
Power Requirements | 115 VAC (100 to 130 VAC), 50/60 Hz |
Weight | 70 lbs (max.) |
Additional Specifications
Performance (as a percent of setpoint)
Accuracy:
From 10 to 100% of Full Scale Flow
Concentration: ± 1.0%
Flow: ± 1.0%
Repeatability: 0.05%
Warm up time: 30 minutes
Mechanical
Gas connections: 1/4" Swagelok (or compatible fitting)
Operating Pressures
Minimum: 10 psig (0.67 Bar)
Recommended: 25 psig (1.68 Bar)
Maximum: 75 psig (5.04 Bar)
Wetted Surfaces
Tubing: Electropolished 316 Stainless Steel
MFC's: Stainless Steel
Seals: Viton® (Optional - Kalrez®, Buna-N, Neoprene, Metal)
Operating temperature: 32° - 122° F (0° - 50° C)
Performance temperature: 59° - 95° F (15° - 35° C)