This product is Discontinued

Malvern Panalytical - Zetasizer µV

Manufactured by  Malvern Panalytical
Follow this Equipment

A highly sensitive batch DLS instrument and a powerful SEC-LS detector, in a single package.

Add the power of advanced light scattering technology to your size exclusion chromatography system (SEC-LS), or use in cuvette mode for large aggregate detection. Uses dynamic light scattering (DLS) to measure molecule size, and static light scattering to determine molecular weight.
Active Questions & AnswersAsk a Question

Particle Size Analyzers Service ProvidersView All (2)

Request Support
Features of Zetasizer µV

Connect the Zetasizer µV to your SEC system using standard chromatography tubing

  • Measure protein size and molecular weight
  • No column calibration
  • Determine protein oligomeric state
  • Measure the polydispersity of each peak
  • Low volume 8µL flow cell
  • Compatible with any SEC system

Simply remove the flow cell and insert a cuvette

  • No calibration required
  • Minimum of 2µL of sample
  • Absolute size measurement
  • Sensitive detection of the presence of aggregates
  • Screen buffers for crystallisation conditions
  • Determine protein melting point or aggregation temperature
  • Measure molecular weight and second virial coefficient
General Specifications
Depth 410 mm
Height170 mm
Width 350 mm
Weight14 kg
Additional Specifications

Temperature control range: 2°C - 90°C +/- 0.1°C*    
Measurement angle: 90°    
Condensation control: Purge facility using dry air    
Standard laser: 60mW, 830nm    
Correlator: Min. sample time 480ns, max. 3600s. 264 channels    
    
Specifications for operation in cuvette mode    
Size range (radius): 0.15nm - 1 micron**    
Minimum sample volume: 2µL    
Sensitivity: 0.1mg/mL of 15kDa protein    
Molecular weight range (estimated from hydrodynamic diameter): 342Da to >106Da     
Molecular weight range (calculated using Debye plot): 980Da to >106Da     
Minimum sample volume: 2µL    
Results output: Molecular size, molecular weight, A2, size distribution    
    
Specifications for operation in chromatography detector mode    
Flow cell volume: 8µL    
Molecular weight range: 106Da     
Results output: Mn, Mw, Mz, Mp, Mw/Mn, Rh, Weight Fraction, Concentration, dn/dc, dA/dc. A2