Dionex - DX-120
Manufactured by Dionex
DX-120 Ion Chromatograph
The Dionex DX-120 Ion Chromatograph performs isocratic ion analysis applications using conductivity detection. The DX-120 is an integrated system, which includes a pump, detector, and injection valve. The chromatography components, including the column(s), Self-Regenerating SuppressorÔ (SRSÔ), and conductivity cell are ordered separately. These components are mounted on the inside of the DX-120 door.
The DX-120 can be controlled locally, using the front panel keypad and display, or remotely, from a PeakNet workstation. The PeakNet workstation consists of a computer with a Dionex DX-LANÔ interface card and PeakNet software (Release 4.30 or higher) installed. Limited remote control is also available using TTL signals.
The DX-120 can be ordered as a single-column system or a dual-column system. The dual-column system allows switching between two sets of columns (column select mode) or between two eluents (eluent select mode). In the eluent select mode, the DX-120 can perform step gradients.
The DX-120 can be controlled locally, using the front panel keypad and display, or remotely, from a PeakNet workstation. The PeakNet workstation consists of a computer with a Dionex DX-LANÔ interface card and PeakNet software (Release 4.30 or higher) installed. Limited remote control is also available using TTL signals.
The DX-120 can be ordered as a single-column system or a dual-column system. The dual-column system allows switching between two sets of columns (column select mode) or between two eluents (eluent select mode). In the eluent select mode, the DX-120 can perform step gradients.
Active Questions & AnswersAsk a Question
Ion Chromatography Service ProvidersView All (21)
Documents & ManualsView All Documents
Features
There are no Features available.
General Specifications
Depth | 16 in |
Height | 23 in |
Maximum Pressure | 4000 psi |
Width | 12 in |
Noise Level | 53 dB (at “A WEIGHING” setting) |
Weight | 65 lbs |
Flow Rate Range | 0.5 to 4.5 ml/min |
Additional Specifications
Control Panel
Display: Liquid crystal display
Buttons: 16 front panel buttons for controlling the display and operating functions
Pump
Type: Single-reciprocating piston with metal-free flow path with PEEK components
Operation Mode: Constant volume
Pulse Damper
Type: Coiled restricted tubing (PEEK), controlled compliance
Detector
Range: 1000 mS, full-scale
Temperature Compensation:
1.5, 1.7, or 1.9% per ºC
Cell Drive: Pulsed, bipolar fixed frequency
Auto Offset: -999 to 999 mS
Local Operation:
Front panel controls and display status of all functions
Remote Operation:
Control of four functions via TTL inputs DX-120 Operator’s Manual