This product is Discontinued

Shimadzu - RF-1501

Manufactured by  Shimadzu
Follow this Equipment

Outstanding Sensitivity - 300 or higher signal-to-noise ratio

Outstanding Sensitivity - 300 or higher signal-to-noise ratio  
The unique RF-1501 optical design provides exceptional sensitivity in an incredibly compact instrument. The bright optical system and noise suppression digital filter allow for high sensitivity measurement. You get a signal-to-noise ratio greater than 300 for the Raman lines of distilled water.  
The scanning speed is selectable in 4 steps, the fastest being 3,700 nm/min.  
And wavelength slewing is possible at ultrahigh speeds reaching 30,000 nm/min.  
  
Pop-in Program Packs to Meet Your Needs  
In seconds, the easy-to-insert Program Packs reconfigure the RF-1501 from the standard quantitative and scanning instrument to a kinetics capable instrument, for example. You can even store your own experimental parameters on convenient Data Packs (IC RAM card).
Active Questions & AnswersAsk a Question

Recent Questions & Answers

Need Equipment Support?

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features

There are no Features available.

General Specifications
Depth 400 mm
Height255 mm
Width 500 mm
Power Requirements 100/120/220/240 V
Peak Power400 VA
Weight23 kg
Additional Specifications

Light Source: Xenon lamp, 150W, Ozone resolving type lamp housing  
Excitation and emission monochromators: Concave, non-aberration, blazed holographic grating, F/2.4, 900 grooves/mm  
Wavelength scale: 220 ~ 900 nm and 0 order  
Measuring wavelength range: 220 ~ 750nm and 0 order as standard  
220 ~ 900nm and 0 order optionally  
Spectral bandwith: Choice of 10 nm and 20 nm in both excitation and emission sides  
Wavelength accuracy: ±5 nm  
Light source compensation: Dynode feedback system with monochromatic light monitoring function  
Sensitivity:] The S/N ratio is 300 or higher for the Raman lines of distilled water (350 nm excitation wavelength, 10 nm spectral bandwidth, and 2-second response)  
Wavelength scanning: 4-step selection of SUPER, FAST, MEDIUM, and SLOW, (about 3,700 nm/min. at SUPER)  
Wavelength slewing speed: About 30,000 nm/min.  
Response: Choice of 0.02, 0.03, 0.1, 0.25, 0.5, 2, and 8 seconds for 98 % of full scale  
Sensitivity selection: 2-step selection of HIGH and LOW (The sensitivity at HIGH is about 50 times that at LOW.)  
Detector: Photomultiplier tube for both excitation and emission sides  
Display: Backlit liquid crystal display, 320 X 200 dots  
Interface: RS-232C interface, and interfaces for printer (Centronics type), autosampler, sipper, and IC card drive  
Operational temperature range: 10 ~ 35 °C  
Operational humidity range: 40~80 % (Below 70% with temperature higher than 30 °C.)

Looking for

Shimadzu - RF-1501?

Shop on LabX.com