Shimadzu - RF-10AXL
RF-10AXL Fluorescence Detector
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Updated byLabWrenchSiteAdmin
Fluorescence Detector Service ProvidersView All (21)
Documents & ManualsView All Documents
Features of RF-10AXL
- The RF-10AXL has a compact, modular design and is used with an SCL-10A system controller to complement any existing HPLC system.
- easily integrates with other Shimadzu LC-10A Series components via fiber optic communication cables for "plug-and-play" convenience in building a customized Shimadzu LC system.
- self diagnostics
- maintenance information (lamp lit hours, lamp energy)
- span calibration functionality
- offers a Certificate of Compliance and Hardware validation standard operation procedures.
General Specifications
Depth | 520 mm |
Height | 205 mm |
Width | 260 mm |
Power Requirements | 100-240 V as ordered, 50/60 Hz |
Peak Power | 350 VA |
Weight | 17 kg |
Additional Specifications
Light source: High Precision Xenon lamp, 150W
Excitation/emission monochromators: Concave, blazed holographic grating monochromators, F2.4
Measuring wavelength range: 200 - 650 nm (200-900nm option)
Spectral bandwidth: 15 nm both in the excitation and emission sides
Wavelength accuracy: +/- 2 nm
Wavelength reproducibility: +/- 0.2 nm
Sensitivity: The S/N ratio is 300 for the Raman line of distilled water (350nm excitation wavelength and 1.5 sec. time constant)
Detector: Photomultiplier
Cell (volume, pressure, material): 12 µL, approx. 20kgf/cm2(2MPa) SUS316, Quartz, PTFE, as standard (2 µL cell and inert cells are available as option)
Wavelength scanning: Possible both for excitation and emission wavelengths.
Time programming: Wavelengths parameters may be programmed in up to 32 steps
Ambient temperature requirements: 4 - 35° C