GE Analytical Instruments - Sievers 500 RL
Continuous Online Monitoring of TOC
Active Questions & AnswersAsk a Question
PRICE REQ
TOC Analyzers Service ProvidersView All (11)
Documents & ManualsView All Documents
Features
There are no Features available.
General Specifications
Depth | 10.8 in |
Measurement Range | 0.03 to 2500 ppb |
On-Line Monitoring | 1 |
Height | 16.5 in |
Width | 19 in |
NDIR | 0 |
Power Requirements | 100–240 ±10% VAC, 50/60 Hz |
Peak Power | 50 W |
Weight | 37.2 lbs |
Portable | 0 |
Additional Specifications
Total Organic Carbon
Measurement Range: 0.03 to 2500 ppb
Accuracy: ± 5% of measurement
Precision: < 1% RSD or 0.03 ppb as C, whichever is greater
pH Range: 5.5 to 8.0 in order to meet accuracy specification
Analysis Modes: On-Line, On-Line Averaged, On-Line Timed
Analysis Time: Continuous mode: 6 minutes
Average and timed modes: 0.5, 1, 4, 8, or 24 hours
Ozone Compatibility3: Maximum 200 ppb as O3
External Flow Rate: Minimum 50 mL/min
Sample Temperature: 1° C to 95° C (34° F to 203° F)
Sample Pressure: Up to 100 psig
Interferences: Insensitive to organic heteroatoms
Calibration Stability: Typically stable for 12 months
Display Readout: 3 significant digits
Conductivity
Raw Conductivity Range4: 0.01 to 35 ΜS/cm
Conductivity Accuracy3,4: ± 0.005 ΜS/cm or ± 1%, whichever is greater
Conductivity Precision: < 0.25% RSD
Maximum Sample Conductivity: 25 ΜS/cm at neutral pH
Instrument
Normal Operating Environment: Intended for indoor use only
Ambient Temperature: 10° C to 40° C (50° F to 104° F)
Maximum Relative Humidity: Up to 95%, noncondensing
Maximum Altitude: 2,300 m (7,546 ft)
Inputs: One isolated binary input
Outputs: Three isolated 4-20mA outputs, one Serial (RS-232), one USB port,
one parallel printer port, four alarm outputs, and one Ethernet port
Installation/Overvoltage: Category II (protects against transients present in Category II power)
Safety Certifications: ETL, CE
Pollution Degree: 2 (normally only non-conductive pollution)
Display: Backlit Quarter-VGA touchscreen display
IP Rating: Environmental enclosure: IP 45