GE Analytical Instruments - Sievers 5310 C On-Line

Manufactured by  GE Analytical Instruments
Follow this Equipment

The Sievers 5310 C On-Line TOC Analyzer delivers high TOC analysis productivity through advanced automation and easy-to-use features.

The Sievers 5310 C On-Line Total Organic Carbon (TOC) Analyzer was designed specifically for the municipal drinking water market for continuous monitoring of raw, finished and plant process waters. It offers superior accuracy and precision across a dynamic operating range of 4 parts per billion (ppb) to 50 parts per million (ppm) TOC.    
    
The 5310 C On-Line Analyzer was engineered for ease of use and cost-effective operation, just like its laboratory model, the Sievers 5310 C Laboratory. It simplifies monitoring and enhances productivity with automated procedures such as calibration, verification, reagent adjustment, and percent removal calculation. The low-maintenance 5310 C needs no external reagents or gas supplies, and offers 12-month calibration stability.    
    
The Analyzer has a dust-resistant and spray-proof IP-45 rated enclosure for demanding water environments. It runs in continuous analysis mode, and can also accept discrete grab samples and standards at any time through the patented Sievers Integrated On-Line Sampling System (iOS System).
Active Questions & AnswersAsk a Question

There are no current Discussions

TOC Analyzers Service ProvidersView All (11)

Request Support
Features of Sievers 5310 C On-Line

On-Line and Grab Sampling     The patented iOS (Integrated On-Line Sampler) lets users test external samples and standards easily without removing the Analyzer from the continuous water source.    

Dual Stream Compatibility and Percent TOC Removal (On-Line)     The 5310 C reports TOC values for two stream and offers the option of automatically calculating a TOC Percent Removal based on user requirements.    

Automatic Determination of Percent TOC Removal (Grab)     If the user selects Grab Mode, the 5310 C On-Line Analyzer can automatically report Percent TOC Removal with the individual TOC results for each vial.    

Grab Mode Alphanumeric Labelling     This feature lets users individually label each grab sample so they can easily recall specific samples.    

Self-Contained Enclosure     The Sievers 5310 C On-Line's rugged IP-45 rated enclosure withstands demanding environments.    

Low Maintenance and Operating Costs     The 5310 C requires just a few hours per year of preventive maintenance. The modular design makes it easy to replace consumables and perform preventive maintenance.    

12-Month Calibration Stability     The 5310 C On-Line Analyzer offers 12-month calibration stability and a variety of single- and multi-point calibration routines.    

Expanded Data Access     The 5310 C features a USB port that allows data transfers from the Analyzer to a USB flash memory drive without interrupting analysis.    

Reliability     The 5310 C On-Line Analyzer delivers superior reliability by using carefully selected materials and components to ensure maximum uptime.

General Specifications
Measurement Range4 to 50000 ppb
NDIR0
On-Line Monitoring1
Portable0
Depth 10.4 in
Height24.6 in
Width 17.8 in
Power Requirements 100–240 ± 10% VAC, 50/60 Hz
Peak Power50 W
Weight37.2 lbs
Additional Specifications

Operating Specifications    
Measurement Range: 4 to 50000 ppb    
Precision: < 1% RSD    
Accuracy: ± 2% or ± 0.5 ppb, whichever is greater    
Display Readout: 3 significant digits    
Sample Type: On-line continuous or discrete grab sample    
Calibration: Typically stable for 12 months    
Analysis Time: 4 minutes    
Sample Temperature: 1–95 °C (34–203 °F)    
Ambient Temperature: 10–40 °C (50–104 °F)    
Sample Pressure: Up to 250 psi    
Required Sample Line Flow Rate: 50–300 mL/min (for on-line mode)    
Instrument Sample Flow Rate: 0.5 mL/min    
    
Analyzer Specifications    
Outputs: 4-20mA output (1); Serial RS-232 output (2); USB port (1); parallel printer port (1)    
Display: Quarter-VGA, color touch-sensitive LCD    
Safety Certifications: ETL, CE see Declaration of Conformity    
Enclosure Rating: IP-45    
    
Consumables    
UV Lamp: 6 months    
Acid Reagent: As needed, typically 3 months with an ICR or 6 months without an ICR    
Oxidizer Reagent: As needed, typically 3-month stability; available in 150- or 300-mL cartridge    
Pump Tubing: 12 months    
Resin Bed: 12 months