ELGA - BIOPURE 60/120
Manufactured by ELGA
Single System Purification for Washer Disinfectors and Endoscope Reprocessors
The BIOPURE 60/120 has an in-built reservoir to ensure minimal space is occupied in the hospital or surgery. Contamination risk to patients is minimized by the use of compliant water maintained through the use of in-line UV oxidation, deionization and molecular ultrafiltration.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Water Purification Service ProvidersView All (6)
Documents & ManualsView All Documents
Features of BIOPURE 60/120
- Delivers up to 10 liters per minute of compliant water.Meets the latest medical standards (EN15883, HTM2030/2031, NHS MESc52, MDA/HIS, ISO 15883 part 4, EN285)
- Guaranteed bacterial specification of <1 CFU/10ml provided by the integrated purification technologies all located inside one single system
- Easy to use monthly disinfection program
- Easy to install under a bench
- Full service and validation support
General Specifications
There are no General Specifications available.
Series Modals
ELGA - BIOPURE 120
[+] View Model SpecificationsFlow Rate | 9 L/min (requirement at 15°C) |
Water Purification Method | RO |
ASTM Type | ASTM Type I |
TOC | 3 ppb |
Depth | 18.9 in |
Height | 32.8 in (without casters) |
Width | 31.1 in |
Power Requirements | 230V ac, 50Hz |
Peak Power | 800 VA |
Weight | 253.5 lbs (operational) |
Additional Model Specifications | Make-up rate @15°C: 120 l/hr Delivery flow rate – maximum: Up to 10 l/min @ 22.5psi (1.5bar) Inorganics (conductivity @25°C): <30 ΜS/cm1 <30 ΜS/cm1 Inorganics (conductivity @25°C) (Optipure Purification Cartridge installed): <5 ΜS/cm Bacteria - typical: <1 CFU/10ml Endotoxin: <0.25 EU/ml Feedwater Requirement Source: Tap water as detailed below Conductivity: <2000 ΜS/cm * Hardness: <250 ppm as CaCO3 Free Chlorine: <1 ppm Cl2 Chloramine: <0.2 ppm Cl2 Silica: <30 ppm SiO2 Fouling Index: <10 FI Iron/manganese: <0.05Fe/Mn ppm Organics: <3 ppm TOC Temperature: 4 – 40°C (Recommended 15 - 25°C) Drain requirements (gravity fall with air gap): 20 l/min Feedwater pressure: 90psi (6bar) maximum, 30psi (2bar) minimum Noise level: <60 dBA |
ELGA - BIOPURE 60
[+] View Model SpecificationsFlow Rate | 9 L/min (requirement at 15°C) |
Water Purification Method | RO |
ASTM Type | ASTM Type I |
TOC | 3 ppb |
Depth | 18.9 in |
Height | 32.8 in (without casters) |
Width | 31.1 in |
Power Requirements | 230V ac, 50Hz |
Peak Power | 800 VA |
Weight | 253.5 lbs (operational) |
Additional Model Specifications | Make-up rate @15°C: 60 l/hr Delivery flow rate – maximum: Up to 10 l/min @ 22.5psi (1.5bar) Inorganics (conductivity @25°C): <30 ΜS/cm1 <30 ΜS/cm1 Inorganics (conductivity @25°C) (Optipure Purification Cartridge installed): <5 ΜS/cm Bacteria - typical: <1 CFU/10ml Endotoxin: <0.25 EU/ml Feedwater Requirement Source: Tap water as detailed below Conductivity: <2000 ΜS/cm * Hardness: <250 ppm as CaCO3 Free Chlorine: <1 ppm Cl2 Chloramine: <0.2 ppm Cl2 Silica: <30 ppm SiO2 Fouling Index: <10 FI Iron/manganese: <0.05Fe/Mn ppm Organics: <3 ppm TOC Temperature: 4 – 40°C (Recommended 15 - 25°C) Drain requirements (gravity fall with air gap): 20 l/min Feedwater pressure: 90psi (6bar) maximum, 30psi (2bar) minimum Noise level: <60 dBA |